Atrandi Single-Microbe DNA Barcoding Kit — protocol quick reference

Source: DGPM02323198001_Single_Microbe_DNA_Barcoding_Kit_V3.pdf Doc. No. DGPM02323198001, Revision V3, 03 November 2023 (V1 2 Oct 2023; V2 13 Oct 2023). Atrandi Biosciences / Droplet Genomics. Research Use Only. Kit P/N CKP-BARK1. 3 experiments, ~10,000 barcoded cells sequenced per experiment.

Evidence marking. 🟢 = verbatim from the source document · 🟡 = derived or inferred (reasoning given; arithmetic verified where possible) · 🔴 = not established in the current declaration; source review remains open. Any claim not marked 🟢 must reach the final page carrying its uncertainty.

This document ends at barcoding. DNA purification and Illumina library prep are in the companion guide DGPM02323206001 — see 02_library_prep.md.


1. Principle

Semi-permeable capsules (SPCs): picolitre aqueous core inside a hydrogel shell. The shell is a size-selective membrane — small reagents (enzymes, detergents, salts) diffuse in and out freely, while genomic DNA and cells are retained. This is what allows a harsh alkaline lysis and multiple buffer exchanges without losing single-cell compartmentalisation, and it is what makes split-pool barcoding possible without any single-cell isolation.

Workflow: encapsulation → fixation → lysis → whole-genome amplification → debranching → end-prep → 4 rounds of split-pool barcode ligation → (companion doc) release, library prep, sequencing.

Total procedure time ≈ 8 h 25 min (3 h 35 min hands-on, 4 h 50 min incubation).

2. Barcode architecture

BarcodePlate columnsn
A1–324
B4–624
C7–924
D10–1224
5'- P7 - Index i7 - Adapter - D - C - B - A - {Fragment} - Adapter - Index i5* - P5 -3'
                             \_______________/
                                cell barcode              (*i5 optional)

3. Kit contents

GroupComponentP/NAmount
Cell lysis2X Lysis Buffer (LB)*CRP-LB13 × 0.4 mL
Neutralization Buffer (NB)CRP-NB13 × 5 mL
WGAdNTP Mix (10 mM each)CRP-DNTP1140 µL
Primer MixCRP-PRM170 µL
WGA PolymeraseCRP-WGP170 µL
WGA EnhancerCRP-WGE125 µL
10X WGA Reaction BufferCRP-WGB1140 µL
Debranching / end-prepDebranching EnzymeCRP-DBE150 µL
10X Debranching BufferCRP-DBB1100 µL
End Prep Enzyme MixCRP-EPE130 µL
End Prep Reaction BufferCRP-EPB170 µL
BarcodingLigation EnzymeCRP-LGE1250 µL
10X Ligation BufferCRP-LGB1750 µL
Barcodes plateCRP-PLT13 plates
Ligation AdapterCRP-LGA110 µL
Buffers50X Wash Buffer (WB)**CRP-WB13 × 1 mL
10X STOP Buffer (SB)**CRP-SB13 × 1 mL
Water, nuclease-free (NFW)CRP-WTR13 × 1 mL

Storage −20 ± 5 °C. \* add 100 µL of 1 M DTT to the tube before use. \\ dilute to 1X in NFW before use.

Not provided: 100% methanol (fixation); 1 M DTT (lysis); 10% Pluronic F-68 / Poloxamer 188 (Thermo 24040032, for WGA); 0.1 M DTT (WGA); 1.5 mL non-stick tubes (e.g. Eppendorf LoBind); SYTO™9 (Thermo S34854, SPC imaging); adhesive plate seals.

Note: the Ligation Adapter (CRP-LGA1) ships in this kit but is only used in the companion library-prep protocol.

4. Protocol

Starting material

Encapsulate on the FLUX SPC Generator or ONYX system. After emulsion breaking and washing, obtain ≥150 µL packed SPCs; working range 150–500 µL. Start with ≥30,000 encapsulated cells to recover 10,000 sequenced genomes. Target occupancy ~0.1 (over-loading causes barcode cross-contamination).

Cell fixation (steps 1–3)

  1. Chill 100% methanol on ice or −20 °C.
  2. To 100 µL SPCs add 900 µL cold methanol drop by drop while agitating.
  3. Store at −20 °C / −80 °C for long-term storage.

Preparation (steps 4–5)

Dilute 50X WB and 10X SB to 1X in NFW.

Cell lysis (steps 6–13)

Skip 6–7 if already fixed after encapsulation. 6–7. Methanol fixation as above; incubate 30 min (or overnight) at ≤ −20 °C.

  1. Wash 3× with 1 mL 1X WB (vortex, 1 min @ 1000 × g, aspirate).
  2. Bring volume to 500 µL with 1X WB.
  3. Prepare fresh alkaline 2X LB by adding 100 µL of 1 M DTT; add 500 µL onto the SPCs, mix.
  4. Rotate 15 min at RT (20–25 °C). Proceed immediately.
  5. Wash 5× with 1 mL Neutralization Buffer (NB).
  6. Wash 5× with 1 mL 1X WB.

Alkaline lysis + DTT. Do not use mechanical lysis with SPCs.

Whole genome amplification (steps 14–20)

ReagentµL
SPCs150
NFW97.5
10X WGA Reaction Buffer37.5
dNTP Mix37.5
Primer mix18.75
0.1 M DTT3.75
10% Pluronic F-683.75
WGA Enhancer7.5
WGA Polymerase18.75
Total375

Aliquot into 5 PCR tubes, 75 µL each.

TempTime
45 °C15 min
65 °C10 min
4 °Chold
  1. Pool into 1.5 mL tube; rinse tubes with 100 µL 1X WB; pellet 1 min @ 1000 × g.
  2. Wash 3× with 1 mL WB.
  3. Recommended: image with 10× SYTO9 to check occupancy of amplified genomes (Figure 9).
  4. Stop point: 4 °C in Stop Buffer up to 7 days; wash 3× with WB before continuing.

🟡 INFERRED. The 45 °C / 15 min + 65 °C profile is consistent with isothermal strand-displacement (MDA-like) amplification. 🔴 Enzyme identity is proprietary — Atrandi state nothing beyond "WGA Polymerase".

DNA debranching (steps 21–24)

ReagentµL
SPCs150
10X Debranching Buffer30
Debranching Enzyme15
NFW105
Total300
  1. 37 °C, 1 h, 1000 rpm on a thermomixer.
  2. Wash 3× with 1 mL 1X WB.
  3. Stop point: 4 °C in WB up to 2 days.

🟡 INFERRED. Presumed to resolve the hyperbranched / cruciform structures produced by strand-displacement WGA into ligatable linear duplex ends. 🔴 The enzyme is not named; a structure-specific resolvase such as T7 endonuclease I would fit, but Atrandi do not say.

DNA end preparation (steps 25–30)

Barcode ligation must follow immediately.

ReagentµL
SPCs150
End Prep Reaction Buffer21
End Prep Enzyme Mix9
Total180

Aliquot into 2 PCR tubes, 90 µL each.

TempTime
20 °C30 min
65 °C30 min
4 °Chold
  1. Pool, rinse with 100 µL 1X WB, pellet 1 min @ 1000 × g.
  2. Wash 3× with 1 mL 1X WB.
  3. Proceed immediately to step 31.

🟡 INFERRED. 20 °C = end repair/blunting, 65 °C = dA-tailing + inactivation — this is the standard NEB-style end-prep profile. Implies 3'-dA ends, hence a T-overhang on the round-A barcode cassette. This inference carries real weight: see 05_barcode_cassette_model.md §2.

Split-and-pool barcoding (steps 31–44)

Thaw the barcode plate at RT, spin 1000 × g for 30 s. Cut the seal only over the wells in use.

Barcode A ligation (steps 31–40) — master mix:

ReagentµL
SPCs150
10X Ligation Buffer60
Ligation Enzyme20
NFW70
Total300
  1. Distribute to 8-strip tubes / 8 wells, 36 µL each (or directly to 24 wells).
  2. 8-channel pipette 10 µL master mix into each well containing 10 µL barcodes A (columns 1–3); mix.
  3. Seal with a fresh adhesive seal.
  4. 22 °C (or RT) for 15 min.
  5. Spin briefly.
  6. Add 40 µL 1X Stop Buffer per well.
  7. 5 min at RT.
  8. Pool into a 1.5 mL tube, pellet 1000 × g 1 min, remove supernatant. Optional: rinse wells with 40 µL SB to maximise recovery.
  9. Wash 5× with 1 mL 1X WB (1 min @ 1000 × g).
  10. Barcode B — repeat 31–40 with columns 4–6.
  11. Barcode C — repeat 31–40 with columns 7–9.
  12. Barcode D — repeat 31–40 with columns 10–12.
  13. Stop point: barcoded sample keeps at 4 °C in 1X Stop Buffer for a few weeks.

Each round: 24-way split, 15 min ligation, stop, pool, wash. 🟡 Reaction volume per well is 20 µL (10 µL mix + 10 µL barcode), quenched with 40 µL SB — arithmetic from the stated volumes.

5. Troubleshooting (Table 1)

IssueCauseSolution
No MDA-positive capsulesCell conc. too lowAdjust before encapsulation
Low yieldInefficient lysisAdjust lysis conditions. Do not use mechanical lysis with SPCs
Unexpected microorganisms in dataContaminationLaminar flow hood, decontaminate surfaces, filter tips; UV the SPC chip and Core Reagent before encapsulation. Do not expose Shell Reagent to light
High barcode cross-contaminationCell conc. too highDilute to occupancy ~0.1
Few SPCs recoveredSPC stickinessNon-stick (LoBind) tubes, pipette carefully

6. Notes and gaps

7. Bounded source pass, 2026-10-17 (round breadth-first-07)

Re-read: preprint v2 (latest; bioRxiv API reports no published version) Methods lines 1322–1393 (encapsulation, lysis R4, MDA, PTA, barcoding, library prep); Bascet trimming lines 897–902; this kit guide V3 steps 1–44 (lines 270–569) and revision table; library guide DGPM02323206001 V3 sections I–VI and sequencing table (lines 161–663).

Model corrections (declaration graph and page):