LamPORE
Defining source
🟢 James et al., “LamPORE: rapid, accurate and highly scalable molecular screening for SARS-CoV-2 infection, based on nanopore sequencing”, medRxiv (2020), doi:10.1101/2020.08.07.20161737.
🟢 Read 2026-10-17: medRxiv v3 (posted 30 Oct 2020; latest of three versions, unpublished per the medRxiv API), obtained from its open medRxiv location (2020.08.07.20161737v3.full.pdf.txt): Methods 1 "Amplification and library preparation" lines 196–256, Methods 2 data analysis lines 257–378, Results "Assay design" lines 380–398, "Barcode demultiplexing" lines 948–968, "FIP barcode optimisation" lines 1002–1022, references lines 1690–1790. Its Supplementary Table 1 (clinical results) was not obtained; it is not a chemistry source. Workflow claims were earlier checked against the later clinical paper, Peto et al., J Clin Microbiol (2021), doi:10.1128/JCM.03271-20 (PMC8315953, JATS Methods "LamPORE", .xml.txt lines 218–220), which used the ONT LamPORE protocol dated 1 July 2020 (version 1, revision 4). That vendor protocol revision is not in the bundle. The JCM paper is a separate clinical study (its own preprint is medRxiv 10.1101/2020.09.18.20195370, cited as reference 20 of the defining preprint, lines 1820–1824, with the JCM title and first author), not a published version of the defining preprint; no published version of the defining preprint is recorded in the source bundle or manifest (checked 2026-10-19).
Molecular workflow
🟢 Four multiplexed RT-LAMP primer sets target SARS-CoV-2 ORF1a, E and N plus human ACTB. One of eight sample barcodes is embedded in each forward inner primer (FIP), so it is copied repeatedly into the branched LAMP products.
🟢 (preprint lines 198–222) Primer sets come from NEB and cited work; short barcodes are added to the FIPs as in LAMP-Seq (Schmid-Burgk) and Yamagishi; primers are IDT HPLC-purified and actin primers are deliberately at lower concentration. One 10x primer pool per FIP barcode; each plate row receives one barcoded FIP mix. Each reaction holds 25 µl 2x LAMP Master Mix (NEB E1700), 5 µl 10x primer pool and 20 µl RNA (50 µl total, computed), incubated at 65 °C for 35 min then 80 °C for 5 min. The JCM summary (65–80 °C, 40 min, line 220) is consistent. Eight of an initial twelve barcoded FIPs were kept after sensitivity screening (lines 1002–1022).
🟢 (preprint lines 222–256) Reactions are pooled by column (12 pools of 8). Each pool gets a 10 µl rapid-barcoding reaction (6.5 µl water, 1 µl pooled LAMP product, 2.5 µl SQK-RBK004 rapid barcode), 30 °C 2 min then 80 °C 2 min. All reactions are pooled into one tube, purified with 0.8x AMPure beads, washed in fresh 80% ethanol and eluted in 15 µl EB. 11 µl eluate receives 1 µl rapid adapter (RAP), 5 min at room temperature, and is sequenced on one MinION flow cell for 1 h. After RBK the fragments have a modal length of about 500 bp, from LAMP strands up to about 5 kb (lines 950–956).
🟢 (JCM lines 218–220) Eight differently FIP-barcoded reactions are pooled by plate column. A rapid transposase-based step adds one of twelve second barcodes, yielding 8 × 12 combinatorial sample identities. All samples are then pooled into a single library without normalization, the pool receives a magnetic bead cleanup, and it is loaded on a MinION flow cell (GridION instrument) and sequenced for 60 min. In that study FIP7 was not used (lower β-actin read counts), so plates held 80 samples plus 2 no-template and 2 synthetic-RNA positive controls; this is study-specific, not a protocol change.
🟢 The preprint resolves two earlier gaps in the JCM Methods: RAP is added after the AMPure cleanup, and the beads are 0.8x AMPure. The adapter attachment chemistry and bases remain commercial and undisclosed, so the motor-adapter input stays an inferred molecular state.
🔴 The commercial FIP-barcode, rapid-barcode, attachment-tag and motor-adapter sequences are not disclosed by the defining sources. The page therefore represents their molecular roles and does not invent bases or one linear consensus for heterogeneous LAMP concatemers.
🟡 The defining preprint (Assay design, lines 386–394; references 10 and 14) cites the ORF1a primer set "AS1" to Rabe & Cepko (medRxiv 10.1101/2020.04.23.2007687, not read) and E1, N2 and actin to the Zhang et al. bioRxiv preprint. Whether Zhang's "As1e" set is identical to the AS1 set used is not checked here.
🟢 The later clinical LamPORE paper (Peto et al., PMC8315953, Methods) says its ORF1a, E and N target sets use previously published primers and cites Zhang et al., doi:10.2144/btn-2020-0078. The published BioTechniques full text was not obtained (manual row), so the primer bases come from Table 1 of the bioRxiv version prints six unbarcoded primers each for As1e (ORF1a), E1, N2 and ACTB. As examples, its F3 primers are As1e CGGTGGACAAATTGTCAC, E1 TGAGTACGAACTTATGTACTCAT, N2 ACCAGGAACTAATCAGACAAG, and ACTB AGTACCCCATCGAGCACG (bioRxiv v1 pdf.txt lines 275–330, checked 2026-10-17; preprint provenance, to be replaced by the published table when obtained).
🟡 The cited set gives defensible target-binding precursor oligos, but the LamPORE product adds a sample barcode into each FIP and the clinical paper does not identify the complete resulting FIP molecules. The cited unbarcoded sequences should not be copied into a LamPORE ordering list as finished FIPs.
Read interpretation
🟢 (JCM line 220; preprint lines 335–378, same thresholds chosen to maximise F1 for 1 h runs) The report counts reads assigned to each target: invalid < 50 classified SARS-CoV-2 + β-actin reads; positive ≥ 50 SARS-CoV-2 reads; inconclusive 20–49; negative < 20.
🟢 (preprint lines 257–333) Classification uses the rapid barcode (guppy_barcoder 4.0.11, SQK-RBK004, mid-strand barcodes detected), the repeatedly copied FIP barcode, and target-derived sequence assigning a read to ORF1a, E, N, ACTB or unclassified. Nanopore acquisition is primer-free and variable-length.
🟢 (preprint lines 962–968) Every LamPORE read carries an ONT rapid barcode at one end; because not every repeat unit carries the FIP barcode, reads containing a LAMP barcode are selected, typically retaining about 70% of reads. This is computational read selection.
🟡 The physical fate of fragments without a rapid-barcode attachment end is not stated (all reactions are pooled before the 0.8x AMPure cleanup), so the model marks them as an unresolved terminal rather than discarded; the LAMP-barcode read selection above concerns a different population and does not assign this fate.