Solid: source-documented · dashed: inferred · dotted gray: guessed earlier chemistry analogue. Guesses are not established ancestry; no link establishes identical chemistry.
No formal ancestor assigned in this review. The scoped inferred predecessor is the most plausible earlier declared design given the inspected mechanism and chronology; no formal descent is asserted.
Long genomic fragments are Tn5-tagged, gap-filled and circularized. After shearing, streptavidin selects short fragments spanning the old circle-closing bond; TruSeq adapters turn those distant genomic ends into an outward-facing read pair.
The drawn duplicated junction is the common product; the vendor also documents single-junction products.
Reaction workflow
Nextera mate-pair junction library
Tagment long genomic DNA (55 °C, 30 min; Zymo column purification). Inputs: Mate Pair Tagment Enzyme biotinylated junction adapter; mate-pair Tn5 transposase
Repair the short single-stranded Tn5 gaps. Inputs: Strand Displacement Enzyme Mix with dNTPs (20 °C, 30 min)
Remove short fragments with AMPure XP. Inputs: AMPure XP beads, about 0.4x (gel-free 40 µl to 100 µl diluted reaction). Other outputs: short tagmentation fragments (< 1500 bp) and reaction mix → discard
Close the distant tagged ends by blunt ligation. Inputs: Circularization Ligase, up to 600 ng DNA in 300 µl, 30 °C overnight. Other outputs: uncircularized long molecules
Shear circular molecules into short fragments. Inputs: Covaris S2/S220, 40 s at 6 °C (~300–1000 bp, 3′/5′ overhangs)
Capture the old circle-junction fragments. Inputs: Dynabeads M-280 Streptavidin (20 °C, 15 min; 4 Bead Wash Buffer and 2 RSB washes). Other outputs: non-junction shear fragments lacking the biotinylated adapter → discard
Repair captured fragment ends. Inputs: End Repair Mix, on bead (30 °C, 30 min)
Add one 3-prime dA to each end. Inputs: A-Tailing Mix, on bead (37 °C, 30 min)
Present adapter and substrate ends. The captured fragment is explicitly dA-tailed before the complementary TruSeq adapter is ligated.
Ligate the TruSeq index adapter. Inputs: TruSeq DNA Adapter Index (one of 12); Ligation Mix, on bead (30 °C, 10 min; Ligation Stop Buffer, bead washes)
Amplify the captured mate-pair library. Inputs: PCR Primer Cocktail (P5-end primer; sequence not printed in the guide); PCR Primer Cocktail (P7-end primer; sequence not printed in the guide); Enhanced PCR Mix, on bead; 10 or 15 cycles (98/60/72 °C)
Clean up libraries with AMPure XP. Inputs: AMPure XP beads on the 45 µl PCR supernatant. Other outputs: short library fragments (< 300 bp) and PCR reagents → discard
Other material outcomes
short tagmentation fragments (< 1500 bp) and reaction mix — discard
the bead supernatant is discarded (guide: removes fragments < 1500 bp)
Molecular structure is not represented in this graph.
digested uncircularized DNA — discard
uncircularized molecules are destroyed before junction recovery
Molecular structure is not represented in this graph.
non-junction shear fragments lacking the biotinylated adapter — discard
supernatant and washes are discarded; the technote notes that incomplete enrichment can carry some over
Molecular structure is not represented in this graph.
short library fragments (< 300 bp) and PCR reagents — discard
Molecular structure is not represented in this graph.
Incomplete enrichment and single-junction products. The data-processing technical note states that non-junction fragments lack the biotinylated adapter and should be removed at enrichment, but incomplete enrichment leaves them as forward-reverse pairs with a < 1,000 bp insert peak. Shearing inside the junction adapter yields a fragment carrying only one original end; single-junction products (one adapter copy) also occur. The drawn duplicated junction is the common product.
Gel-plus branch not drawn. The gel-free path (1 µg input, 2–15 kb, median 2.5–4 kb) is drawn. The gel-plus path uses 4 µg input, a 100 µl bead addition to the 200 µl reaction, then Pippin Prep or agarose size selection with Zymo recovery before circularization; PCR is 10 or 15 cycles depending on recovered amount and size.
Final sequencing library
Single-index TruSeq library containing a selected duplicated mate-pair junction. The two genomic pieces flanking it were distant ends of one long input fragment.
Sequencing primers
Standard TruSeq Read 1, six-base Index 1 and Read 2 primers bind the final library. The genomic mates map outward because circularization inverted their relationship.
Sequences come from lib/; the landing site and the first bases of each read are computed from this page's final library, so they cannot drift from it.
anneals to the top strand over dA junction, Index 1 / Read 2 arm
rrrrrrrrrrrr… (from right distant end)
Read content and handling
Content and cycle positions are derived from the final construct, primer direction, molecular-feature annotations and declared run lengths. ‘No’ means the read is auxiliary or absent, not that its information is unimportant.
Read
Cycles
Content
Include in sequence analysis?
Read 1
not fixed
left distant end (from cycle 1; molecule-dependent length); short molecules may read through into the opposite library end
Yes — include in sequence analysis
Index 1 (i7)
not fixed
sample index (cycles 1–6)
No — use for sample demultiplexing
Index 2 (i5)
—
not acquired
No — not sequenced
Read 2
not fixed
right distant end (from cycle 1; molecule-dependent length); short molecules may read through into the opposite library end