Olink Explore / NGS-PEA — source notes

Defining source: Wik et al., Molecular & Cellular Proteomics (2021), doi:10.1016/j.mcpro.2021.100168.

Evidence read

Molecular workflow

🟢 Each protein assay uses a matched pair of antibody-conjugated oligos. One arm contains P5 and the Read 1 sequencing-primer site; the other contains a common reverse preamplification site. Both carry assay-specific barcodes and complementary hybridization arms. Co-binding to the same protein raises their effective local concentration, allowing the arms to hybridize.

🟢 The PCR1 mix (article) or extension mix (patent worked example) is added to each incubation reaction or plate, with no wash described between the immunoreaction and extension (JATS lines 164–168; patent text lines 4128–4135, 4994–4996). The article says only that proximal matched pairs generate detectable amplicons (Fig. 1A legend, line 152). 🟔 Free and non-proximal probes are therefore inferred to stay in the reaction; the model leaves their later physical fate unresolved rather than calling them discarded.

🟢 Samples are used undiluted and at 1:10, 1:100 and 1:1000; 0.2 µl sample is mixed with 0.6 µl block incubation mix (80 µl incubation solution + 10 µl forward + 10 µl reverse probes) and incubated at 4 °C for 16–24 h (PMC8633680 JATS lines 162–164).

🟢 PCR1 mix (Olink PCR1 solution, PCR1 enhancer, PCR1 Enzyme and IDT universal primers) is added; PCR1 begins with 20 minutes at 50 °C for polymerase extension across the paired arms, then 95 °C 5 min and 25 cycles (95 °C 30 s, 54 °C 1 min, 60 °C 1 min). Products from four concentration blocks are pooled for each sample. PCR2 uses the same forward primer and one of 96 reverse primers to add an inline sample index and P7 during ten cycles (95 °C 30 s, 68 °C 1 min). All PCR2 samples of a panel are pooled into one of four libraries, and each library (50 µl) is purified with 80 µl AMPure XP beads to remove primers, primer dimers and dNTPs, then sized on a Bioanalyzer HS chip (JATS lines 168–172). The patent's worked example names Pwo polymerase for extension/PCR1 (line 4135); the article names only "PCR1 Enzyme", so the model does not assign the enzyme identity.

🟢/🟔 Apparent wording discrepancy, shown on the schematic: the article calls the PCR1 forward primer "complementary to Illumina's P5 adapter sequence" (JATS line 168), whereas the patent says the forward common primer "comprises" the P5 adapter (SEQ ID NO: 1; lines 4135–4171). The model uses the P5-sense identity, which the article's own probe design (forward probe carries P5 and Rd1SP) requires.

🟢 The finished amplicon is 148 bp. Its published element order is P5, Rd1SP, forward assay barcode, hybridization-derived junction, reverse assay barcode, sample index and P7.

🟢 WO2021191442A1 SEQ ID 1 is the 20-nt P5 adapter portion AATGATACGGCGACCACCGA; the patent states that the PCR1 forward common primer contains it and the same primer is reused in PCR2. SEQ ID 2 is the 24-nt P7 adapter portion CAAGCAGAAGACGGCATACGAGAT; each PCR2 reverse primer contains it plus a sample-index barcode. These are published portions, not the complete primer sequences.

🟔 The rest of the exact standard Illumina P5/Rd1SP end shown on the schematic comes from Illumina's adapter documentation. The combined end is 58 nt. Together with the patent's 24-nt P7 adapter, the published 148-bp total leaves exactly 66 nt between Rd1SP and P7. This agrees with the published 66-cycle read covering the inline assay and sample identifiers; it is a consistency check, not proof of the bases, and the article does not state where Read 1 starts or ends. The individual lengths within those 66 nt have not yet been verified from an original oligo table.

šŸ”“ The original article, its four supplements and the checked patent do not supply the complete probe oligos, universal PCR primers, assay-barcode whitelist, hybridization arms or sample-index sequences in the inspected sections. This is a provisional source gap: the software archive is password protected, the checked preprocessing guide is from 2026, and third-party example counts have not yet received an independent second source check. The schematic therefore shows these regions by identity and order without inventing bases or individual lengths. No raw-read accession was found in the article's data availability statement; this absence claim also remains under review.

Additional public material checked

Sequencing

🟢 Each of four panel libraries is loaded in one NovaSeq 6000 XP lane. A single 66-cycle Read 1 covers both assay barcodes and the inline sample index. There are no separate index reads: exact barcode combinations are counted to estimate protein abundance after control normalization.

Review trail