PrimeFlow RNA Assay
1. What it is
A commercial branched-DNA fluorescence in situ hybridization assay for measuring selected RNA targets in individual suspension cells by flow cytometry, optionally together with surface and intracellular proteins.
2. Sources and evidence
- 🟢 Thermo Fisher PrimeFlow RNA Assay Kit User Guide, MAN0019788 Rev B.0, dated 22 February 2022.
- 🟢 Porichis F, Hart MG, Griesbeck M, et al. High-throughput detection of miRNAs and gene-specific mRNA at the single-cell level by flow cytometry. Nature Communications 5:5641 (2014). doi:10.1038/ncomms6641.
- 🟢 Supplementary Table 1 gives the target binding intervals and target-region sequences used for the IL-2, IL-21, IFN-γ, IDO, HMBS and miRNA probe sets in the paper.
- 🟢 The article's Methods says that probe sequences are detailed in Supplementary Table 1. The table itself is headed “Sequence of probe sets used in this study,” but its columns give target genes, binding coordinates and target-region bases, not the complete manufactured probes. This schematic derives complementary target-binding arms from those bases; the universal tail remains unresolved.
- 🟢 The manual specifies 20–40 adjacent target-probe pairs, then preamplifier, amplifier and fluorescent label-probe hybridizations. One complete tree has 400 label sites and the full target set gives 8,000–16,000-fold theoretical signal amplification.
- 🟢 The current assay supports up to four targets using Types 1, 4, 6 and 10, followed by standard flow-cytometric analysis.
- 🟢 The 2014 Porichis article says its then-available reagents supported three fluorochromes (Alexa 488, 647 and 750); the later 2022 manual lists Types 1, 4, 6 and 10. The schematic follows the 2022 commercial kit and does not assign Type 10 to the 2014 experiment.
- 🟢 Guide Day 1 steps 4–19 (p.15–16): surface stain 30 min 2–8 °C → Fixation Buffer 1 30 min 2–8 °C → two permeabilization washes with RNase inhibitors → optional intracellular stain → Fixation Buffer 2 60 min room temperature → washes. Day 1 target hybridization: probes 1/20, 2 h at 40 °C, washes, overnight 2–8 °C. Day 2 (p.18): PreAmp Mix 1.5 h at 40 °C, three washes; Amp Mix 1.5 h at 40 °C, two washes; Label Probes 1/100, 1 h at 40 °C, two washes. Porichis Methods report the same 1:20/2 h, 1.5 h, 1.5 h and 1:100/1 h hybridizations at 40 °C after 4% PFA fixation, cold-methanol permeabilization and a second 4% PFA fixation.
- 🟢 Published inconsistency (2022 guide): the product description (p.5) says “three different amplification structures” allow up to four RNA targets; the assay guidelines (p.9) and workflow figure list four probe types (1, 4, 6, 10) with distinct dyes. The page keeps four types and makes no claim about the number of distinct structures. The positive-control table header also prints “Type 2” over VA4/VB4 (Type 4) catalogue numbers.
- 🟢 Only 400 label sites per complete tree is stated; no amplifier-per-preamplifier or label-per-amplifier split is given, so the model carries the total only (400 per tree) and the page draws representative arms without per-level counts.
- 🟡 Those published target regions determine the reverse-complementary target-binding arms. They are not complete orderable probe sequences: the universal tails attached to the arms are absent from the table.
- 🔴 The universal probe tails (including their lengths and docking bases) and the preamplifier, amplifier and fluorescent label-probe sequences are not given in the reviewed supplement or 2022 vendor guide. The original article has been checked twice; the original vendor-hosted 2014 FlowRNA manual and linked Player 2001 predecessor remain unresolved, so an absence claim is not accepted.
3. Molecular path
After optional antibody staining, suspension cells are fixed and permeabilized. Gene-specific probe pairs bind adjacent regions of each target RNA. Preamplifier and amplifier hybridizations build a branched scaffold, fluorescent label probes occupy its terminal sites, and a flow cytometer measures fluorescence in each retained cell. The target RNA is never amplified.
4. Public sequence example
For human IL-2 (NM_021803), the first adjacent target regions are bases 2–24, CCTTGGTCTCGTTTTCACTTCAG, and bases 25–52, TCTCATAAGTACCAACAGTAGAGCTAGA. The schematic derives the two complementary probe arms from these published target sequences and leaves each proprietary universal tail visibly unresolved.