No curated ancestry links yet. This does not establish independent origins.
No formal ancestor assigned in this review. Earlier RNA in-situ hybridization/branched-DNA amplification is relevant. dRNAseq/CAGEscan suggestions do not capture the paired-probe signal amplification chemistry, so they are rejected.
Adjacent double-Z probes recognize RNA in fixed cells or tissue and nucleate a branched-DNA signal tree for fluorescent or chromogenic microscopy. The RNA itself is not amplified.
Sequence boundary. The defining paper withholds the commercial target-probe, preamplifier, amplifier and label-probe sequences. The older core patent nevertheless discloses exact GAPD double-Z probes and a matching preamplifier example; those public sequences are shown below but are not asserted to be the current kit reagents.
(1) Prepare the specimen
Wang 2012 Methods: cultured cells on slides are fixed in 4% formaldehyde for 60 min and digested with protease (2.5 µg/mL) at 23–25 °C. 5-µm FFPE sections are deparaffinized in xylene, taken through an ethanol series, boiled (100–103 °C) in citrate buffer pH 6 for 15 min (printed as "10 nmol/L"; a nanomolar retrieval buffer is implausible and 10 mmol/L is likely intended, but the model does not assert the corrected unit), rinsed and digested with 10 µg/mL protease at 40 °C for 30 min. Current commercial kits may use different pretreatment reagents.
(2) Hybridize adjacent target-probe pairs
Each probe has an 18–25-nt target-complementary region, spacer and 14-nt Z tail. Only two probes bound adjacently create the complete 28-nt preamplifier site. The paper typically targets ~1 kb with 20 pairs and used 10–20 pairs; three pairs already gave visible HRP/DAB signal. Target probes hybridize at 40 °C for 3 h in buffer A (6× SSC, 25% formamide, 0.2% lithium dodecyl sulfate, blocking reagents), followed by three room-temperature washes (0.1× SSC, 0.03% LDS).
Patent-disclosed double-Z sequence example
US7709198B2 reports these oligos for a human GAPD experiment. They establish one exact implementation of the cooperative 28-base junction, not the identity of present-day commercial RNAscope reagents.
example preamplifier (SEQ ID NO:1): 5'- AGGCATAGGACCCGTGTCTTTTTTTTTTTAGGCATAGGACCCGTGTCTTTTTTATGCTTTGACTCAGAAAACGGTAACTTC -3'
The docking site is computed and checked as the reverse complement of the two 14-nt tails. The tails reside on separate target probes; the shared line only makes their cooperative duplex with the preamplifier legible.
Nucleotide-resolved double-Z fold
The patent example with every disclosed base shown. Target RNA is kept straight; five-thymidine spacers are flexible joints, and the two 14-nt Z tails jointly form the preamplifier docking duplex.
(3) Assemble the branched-DNA signal tree
Wang 2012 Methods: sequential 40 °C hybridizations, each followed by three room-temperature washes (0.1× SSC, 0.03% LDS): preamplifier 2 nM in buffer B (20% formamide, 5× SSC, 0.3% LDS, 10% dextran sulfate, blocking reagents) for 30 min; amplifier 2 nM in buffer B for 15 min; label probe 2 nM in buffer C (5× SSC, 0.3% LDS, blocking reagents) for 15 min. Multiplex assays use equimolar reagents of each amplification system.
Five representative amplifier arms are drawn without crossings; arm labels give the modelled per-level multiplicities. Wang et al. 2012 (Methods, RNAscope Design Strategy) state 20 amplifier binding sites per preamplifier and 20 label-probe binding sites per amplifier. The modelled tree carries 400 label-probe sites, and the documented target-probe range gives 4,000–8,000 theoretical label sites per target RNA.
(4) Detect the signal
Wang 2012 Methods: label probes were conjugated to Alexa Fluor 488, 546, 647 or 750 for epifluorescence, or to HRP for DAB chromogen with hematoxylin counterstain and bright-field microscopy (the design section also names alkaline phosphatase/Fast Red). Spots are counted per cell; UBC and bacterial dapB probe sets serve as positive and negative controls.
Measurement, not sequencing
RNAscope produces spatial microscopy signal, not a sequencing library. There are therefore no flow-cell adapters, sample indices, sequencing primers or nucleotide reads to draw.
Read content and handling
This is a branched-DNA imaging/flow-cytometry assay, not an NGS library.