No formal ancestor assigned in this review. The vendor HS2 workflow could descend from older SureSelect hybrid capture, but no exact earlier version/date/source identity has been assigned from this guide. Do not substitute a generic indexed DNA library as a confirmed parent.
A dual-indexed, optionally duplex-MBC-tagged Illumina library is pooled, hybridized to biotinylated RNA baits, recovered on streptavidin beads and amplified after capture.
The B0 guide (G9957-90000) publishes the MBC read layout (five read-start bases at each end, of which one suggested approach keeps the first three as MBC and discards two) and 384 paired eight-base indexes, but not the HS2 adaptor, indexing-primer or post-capture primer bases. Canonical TruSeq arms in the final construct are inferred for orientation and primer landing, not established kit oligos; the selected index pair is unknown. The MBC-free adaptor branch is not drawn.
Reaction workflow
SureSelect XT HS2 captured library (MBC-tagged)
Fragment genomic DNA. Inputs: Covaris shearing (duration by read length) or SureSelect enzymatic fragmentation, 50 µL
Repair fragment ends and add 3-prime dA (20 °C 15 min, 72 °C 15 min). Inputs: End Repair-A Tailing Buffer and Enzyme Mix
Present adapter and substrate ends. Agilent specifies ligation after dA-tailing but does not publish the HS2 adapter terminal bases.
Ligate the HS2 adapter mix (20 °C 30 min). Inputs: XT HS2 Adaptor Oligo Mix (MBC-tagged; the MBC-Free Adaptor Oligo Mix branch is not drawn); T4 DNA Ligase in Ligation Buffer
AMPure XP clean-up (80 µL beads, elute 35 µL). Inputs: AMPure XP beads. Other outputs: AMPure supernatant (unligated adaptor, reaction components) → discard
Index the library by PCR (98 °C 2 min; 8–14 cycles by input amount and quality; 72 °C 5 min), then AMPure (50 µL). Inputs: SureSelect XT HS2 Index Primer Pair, P5 member (8-nt i5); SureSelect XT HS2 Index Primer Pair, P7 member (8-nt i7); Herculase II Fusion DNA Polymerase
Pool indexed libraries before capture. Additional template inputs: other uniquely dual-indexed HS2 libraries
Hybridize target intervals to the probe, 65 °C or 67.5 °C for 16–24 h. Inputs: SureSelect XT HS2 Blocker Mix (95 °C 5 min, then held at the hybridization temperature); target-specific SureSelect probe (biotinylated RNA baits) with RNase Block in hybridization buffer
Capture bait-bound molecules (30 min, room temperature) and wash off unbound library. Inputs: SureSelect Streptavidin Beads; Wash Buffer 1, then 6 × Wash Buffer 2 at 70 °C. Other outputs: unbound off-target library molecules → discard
Amplify the captured library on the beads (98 °C 2 min; 10–16 cycles by probe size; 72 °C 5 min), then remove the beads on a magnet. Inputs: SureSelect Post-Capture Primer Mix: P5-end primer (bases not published); SureSelect Post-Capture Primer Mix: P7-end primer (bases not published); Herculase II Fusion DNA Polymerase. Other outputs: streptavidin beads with bound probe → discard
AMPure XP clean-up of the final pool (50 µL beads). Inputs: AMPure XP beads. Other outputs: AMPure supernatant → discard
Molecular structure is not represented in this graph.
unbound off-target library molecules — discard
unbound molecules are removed with the supernatant and stringent washes
Molecular structure is not represented in this graph.
streptavidin beads with bound probe — discard
beads removed on the magnetic stand; the supernatant is kept
Molecular structure is not represented in this graph.
AMPure supernatant — discard
bead supernatant removed
Molecular structure is not represented in this graph.
Final sequencing library
INFERRED — canonical TruSeq arms place the manual's five read-start bases at each insert end (3-nt MBC + 2 discarded bases) and the eight-base index windows for orientation only; Agilent's exact arm sequences are unresolved.
Sequencing primers
The manual specifies standard Illumina paired-end and dual-index primers with no custom sequencing primers, and 8-bp i7 and i5 index reads. Read 1/Read 2 length is user-chosen: Table 46 lists 2 × 100 or 2 × 150 bp runs and Table 47 gives 151 + 8 + 8 + 151 cycles as an example (one extra cycle per read).
Sequences come from lib/; the landing site and the first bases of each read are computed from this page's final library, so they cannot drift from it.
anneals to the top strand over Read 2 site terminal base, Index 1 / Read 2 arm
vvvssnnnnnnn… (from right MBC)
Read content and handling
Content and cycle positions are derived from the final construct, primer direction, molecular-feature annotations and declared run lengths. ‘No’ means the read is auxiliary or absent, not that its information is unimportant.
Read
Cycles
Content
Include in sequence analysis?
Read 1
not fixed
UMI (cycles 1–3); left non-MBC read-start bases (cycles 4–5); captured target insert (from cycle 6; molecule-dependent length); short molecules may read through into the opposite library end
Yes — after extracting identifiers
Index 1 (i7)
8
sample index (cycles 1–8)
No — use for sample demultiplexing
Index 2 (i5)
8
sample index (cycles 1–8)
No — use for sample demultiplexing
Read 2
not fixed
UMI (cycles 1–3); right non-MBC read-start bases (cycles 4–5); captured target insert (from cycle 6; molecule-dependent length); short molecules may read through into the opposite library end