TIME-seq

Griffin et al., “TIME-seq reduces time and cost of DNA methylation measurement for epigenetic clock construction,” Nature Aging (2024), doi:10.1038/s43587-023-00555-2.

Sources read

Construct-changing steps

  1. 🟢 Each 100-ng genomic-DNA sample is tagmented for 15 min at 55 °C with a uniquely indexed TIME-seq transposome. Adaptor B carries a five-base internal sample barcode; methylated adaptor A and indexed adaptor B are separately annealed to the 5′-phosphorylated reverse mosaic-end oligo when transposomes are assembled. 100 ng unmethylated lambda DNA is tagmented with each pool as a bisulfite-conversion control. STOP buffer (guanidine thiocyanate) is added and incubated 5 min at 55 °C.
  2. 🟢 Samples are combined by pool with Zymo DNA binding buffer and purified on a Clean & Concentrator-25 column. Klenow fragment (3′→5′ exo−) in NEBuffer 2 performs 30 min of end repair at 37 °C using a dNTP mix in which 5-methyl-dCTP replaces dCTP; a second column and SpeedVac concentrate the pool to 5 µL. 🟡 Methyl-dCTP protects newly copied adapter/gap cytosines against bisulfite conversion.
  3. 🟢 DNA with Cot-1 DNA and the A/B blocking oligos is denatured 5 min at 95 °C; biotinylated RNA baits (750 ng) hybridize to target loci for 4 h at 65 °C. RNA:DNA hybrids are captured on streptavidin beads, stringently washed and eluted as single-stranded DNA with NaOH.
  4. 🟢 Captured DNA is converted with the EpiTect Fast Bisulfite kit, then amplified for 19 cycles with Q5U and indexed P5/P7 primers, then cleaned with 1.8× SPRI beads. The PCR indices distinguish pools; the five-base transposome index distinguishes samples within a pool.

Oligos and read layout

🟢 Supplementary Table 5 publishes 64 indexed B adaptors (TS301–TS364). The schematic uses TS301 (AAAGA) as one explicit representative and N501/N701 as representatives of the published N501–N506 (P5/i5) and N701–N704 (P7/i7) PCR primers. 🟢 Each N70x primer is P7, the 8-nt i7 pool index, the 23-nt custom sequencing handle AAGCAGTGGTATCAACGCAGATC and the 14-nt B-adaptor prefix TGGGTGGAGGGTGG; the custom Read 2 primer is handle + prefix and the custom Index 1 primer is its reverse complement (Table 5, sheets “Indexed PCR Primers” and “Sequencing Primers”). It preserves /5Phos/ and /iMe-dC/ exactly as written in the source.

🟡 Published discrepancy (computed from Supplementary Table 5, sheet “Hyb Blocking Oligos”, checked in the original workbook): the “Adaptor A Blocking Oligo” CTGTCTCTTATACACATCTCCGAGCCCACGAGAC is the reverse complement of the Nextera Tn5ME-B arm GTCTCGTGGGCTCGG plus the mosaic end, not of TIME-seq methylated adaptor A (arm TCGTCGGCAGCGTC); only its 19-nt mosaic-end portion pairs with adaptor A. The B blocker is the exact reverse complement of adaptor B (HHHHH opposite DDDDD). The model keeps the published sequence as a hybridization input and draws no blocker duplex; which sequence was intended is unresolved. Minor table inconsistencies, also shown on the page: the P5 structure row has a 7-N i5 placeholder under an “(8-nt)” header while all listed indexes have 8 nt, and the adaptor B structure row labels the 14-nt prefix “Adaptor A”.

🟢 TIME-seq requires its published custom Read 2 and Index Read 1 primers. Read 1 and Index 2 use the corresponding Nextera sites. The economical MiSeq/NextSeq program is Read 1 145–153 cycles, i7 8 cycles, optional i5 8 cycles and Read 2 5 cycles. Thus the short Read 2 reports the internal Tn5 sample barcode.

Derived observations and open questions

Scope

The page documents library construction and sequencing geometry, not epigenetic-clock training, benchmark performance or bait-design results. Bait sequences are target-set specific and are not reproduced as “key oligos”; the adapter-blocking oligos are.