Calibration and quality control
Decide whether a signal can support a concentration
What relation turns signal into concentration?
![Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.]()
Known final concentrations give measured responses.
Fit the measured response direction
![Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.]()
A=a+bc+e.
Retrieve lm
fit <- lm(absorbance ~ standard_mg_l, data = standards)
coef(fit)
Interpret the slope with units
![Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.]()
Slope ≈ 0.198799 L/mg Fe.
Retain the estimated intercept
![Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.]()
Intercept ≈ 0.003874.
Bracket the intended measurement range
![Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.]()
These standards cover 0–4 mg Fe/L final.
Residuals are observations minus fitted responses
![Iron calibration residuals scatter around zero without an obvious curved pattern in the selected range.]()
What pattern would challenge a straight line?
A high R² can hide systematic error
![A deliberately curved response leaves systematic patterned residuals after a straight-line fit.]()
The second dataset is deliberately curved.
Investigate before excluding a standard
![A deliberately curved response leaves systematic patterned residuals after a straight-line fit.]()
A point is not wrong merely because it worsens the fit.
A narrower range needs evidence
![A deliberately curved response leaves systematic patterned residuals after a straight-line fit.]()
Which standards and checks would validate the proposed range?
Changing variance is another model concern
![Iron calibration residuals scatter around zero without an obvious curved pattern in the selected range.]()
Ordinary least squares assumes the specified response-error structure.
Invert only after checking applicability
![Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.]()
\hat c=(\bar A-a)/b.
A concentration estimate is not the original sample result
![A ten millilitre aliquot is made to fifty millilitres total; multiply the final concentration by five to return to the original water basis.]()
The current unit is mg Fe/L in the final coloured solution.
Three materials ask different QC questions
![A blank, independent reference check and a sample spike examine different parts of the measurement.]()
Which would reveal introduced contamination?
Independence matters for a check standard
![A blank, independent reference check and a sample spike examine different parts of the measurement.]()
The same wrong stock can make standards and checks agree.
Acceptance criteria precede the results
What action follows a failure?
A weak signal can be detectable but poorly quantified
![Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.]()
Distinguishing a signal from blanks is a different requirement from precise quantification.
Estimate blank variation from independent blanks
![Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.]()
Mean = 0.0100; SD = 0.001348 signal units.
Keep a heuristic in its proper role
![Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.]()
3s_{blank}/b=0.02023 mg/L; 10s_{blank}/b=0.06742 mg/L.
A signal threshold includes the blank level
![Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.]()
Threshold here: mean blank plus three SD.
Uncertainty remains after calibration inversion
![Averaging repeated readings reduces independent reading noise; it does not remove a common stock or preparation effect.]()
Reading noise and uncertain slope/intercept both contribute.
Use the supplied helper with its assumptions
source("R/inverse-calibration.R")
inverse_calibration(fit, unknown_absorbance)
Interpret the interval on the correct scale
Final concentration ≈ 2.3981 mg Fe/L; approximate 95% interval 2.3886–2.4077.
Transfer: decide on three failures
![A blank, independent reference check and a sample spike examine different parts of the measurement.]()
Outside range; failed independent check; below quantification requirement.
Practise the chemical argument
Open the exercise.
Save the calculation and a chemical interpretation in your reproducible report.
Interactive experiment · Predict, change, explain.