Calibration and quality control

Decide whether a signal can support a concentration

Analytical Chemistry

What relation turns signal into concentration?

Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.

Known final concentrations give measured responses.

Fit the measured response direction

Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.

A=a+bc+e.

Retrieve lm

fit <- lm(absorbance ~ standard_mg_l, data = standards)
coef(fit)

Interpret the slope with units

Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.

Slope ≈ 0.198799 L/mg Fe.

Retain the estimated intercept

Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.

Intercept ≈ 0.003874.

Bracket the intended measurement range

Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.

These standards cover 0–4 mg Fe/L final.

Residuals are observations minus fitted responses

Iron calibration residuals scatter around zero without an obvious curved pattern in the selected range.

What pattern would challenge a straight line?

A high R² can hide systematic error

A deliberately curved response leaves systematic patterned residuals after a straight-line fit.

The second dataset is deliberately curved.

Investigate before excluding a standard

A deliberately curved response leaves systematic patterned residuals after a straight-line fit.

A point is not wrong merely because it worsens the fit.

A narrower range needs evidence

A deliberately curved response leaves systematic patterned residuals after a straight-line fit.

Which standards and checks would validate the proposed range?

Changing variance is another model concern

Iron calibration residuals scatter around zero without an obvious curved pattern in the selected range.

Ordinary least squares assumes the specified response-error structure.

Invert only after checking applicability

Eighteen iron standards from zero to four mg/L define an absorbance line with a retained intercept.

\hat c=(\bar A-a)/b.

A concentration estimate is not the original sample result

A ten millilitre aliquot is made to fifty millilitres total; multiply the final concentration by five to return to the original water basis.

The current unit is mg Fe/L in the final coloured solution.

Three materials ask different QC questions

A blank, independent reference check and a sample spike examine different parts of the measurement.

Which would reveal introduced contamination?

Independence matters for a check standard

A blank, independent reference check and a sample spike examine different parts of the measurement.

The same wrong stock can make standards and checks agree.

Acceptance criteria precede the results

What action follows a failure?

A weak signal can be detectable but poorly quantified

Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.

Distinguishing a signal from blanks is a different requirement from precise quantification.

Estimate blank variation from independent blanks

Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.

Mean = 0.0100; SD = 0.001348 signal units.

Keep a heuristic in its proper role

Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.

3s_{blank}/b=0.02023 mg/L; 10s_{blank}/b=0.06742 mg/L.

A signal threshold includes the blank level

Twelve independent blank responses cluster near 0.010; a dashed line marks mean plus three standard deviations.

Threshold here: mean blank plus three SD.

Uncertainty remains after calibration inversion

Averaging repeated readings reduces independent reading noise; it does not remove a common stock or preparation effect.

Reading noise and uncertain slope/intercept both contribute.

Use the supplied helper with its assumptions

source("R/inverse-calibration.R")
inverse_calibration(fit, unknown_absorbance)

Interpret the interval on the correct scale

Final concentration ≈ 2.3981 mg Fe/L; approximate 95% interval 2.3886–2.4077.

Transfer: decide on three failures

A blank, independent reference check and a sample spike examine different parts of the measurement.

Outside range; failed independent check; below quantification requirement.

Practise the chemical argument

Open the exercise.

Save the calculation and a chemical interpretation in your reproducible report.

Interactive experiment · Predict, change, explain.