Electrodes and potentiometry

Activity, calibration and the complete cell

Analytical Chemistry

About this lecture

  • Explain why the complete cell and activity conditions matter.
  • Fit and invert a logarithmic fluoride calibration in R.
  • Separate a shared drift failure from sample-specific matrix mismatch.

A voltage belongs to a cell

An indicator electrode and reference electrode contact the same sample and connect to a high-input-impedance meter. There is negligible net measurement current.

Which two functions are required?

The indicator responds; the reference anchors

An indicator electrode and reference electrode contact the same sample and connect to a high-input-impedance meter. There is negligible net measurement current.

The measured potential also includes the reference and junction.

Why use a high-input-impedance meter?

An indicator electrode and reference electrode contact the same sample and connect to a high-input-impedance meter. There is negligible net measurement current.

Read potential with negligible net measurement current.

Activity gives the thermodynamic response

Activity is dimensionless and related to concentration by an activity coefficient and standard concentration. Matching ionic strength supports calibration transfer.

a_i=\gamma_i c_i/c^\circ.

Write the logarithmic relationship

E=K+(2.303RT/zF)\log_{10} a_i.

Anion and cation slopes need a convention

Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.

For this fluoride example, increasing concentration lowers the potential.

Match conditions before using concentration

Activity is dimensionless and related to concentration by an activity coefficient and standard concentration. Matching ionic strength supports calibration transfer.

Ionic-strength adjustment helps; it does not remove every interference.

pH is already logarithmic

Illustrative pH potential decreases by 59.16 mV per pH unit at 25 degrees Celsius; the cell sets its intercept.

pH=-\log_{10}a_{H^+}.

Buffers establish the pH calibration

Illustrative pH potential decreases by 59.16 mV per pH unit at 25 degrees Celsius; the cell sets its intercept.

Use appropriate buffers and measurement temperature.

Inspect the fluoride standards

The same fluoride measurements form a curved relationship on an untransformed concentration axis.

Range here: 10⁻⁵ to 10⁻³ mol/L at matched conditions.

Transform the explanatory quantity

mutate(log_c = log10(fluoride_mol_l))

The correct axes reveal the model

Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.

The same observations are nearly linear against log concentration.

Read the slope per decade

Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.

Fitted slope = −59.10 mV per decade.

Invert in two steps

First (E-a)/b; then c=10^{(E-a)/b} mol/L.

Report the accepted early sample

Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.

121.2 mV → 2.00 × 10⁻⁴ mol/L.

Check the calibration at a known concentration

The early fluoride check has error 0.12 mV and passes; the later check has error 6.02 mV and fails the two-mV tolerance.

The check is 1.00 × 10⁻⁴ mol/L; exercise tolerance ±2.0 mV.

The early block passes

The early fluoride check has error 0.12 mV and passes; the later check has error 6.02 mV and fails the two-mV tolerance.

Observed 139.1 mV; error +0.12 mV.

The later block fails

The early fluoride check has error 0.12 mV and passes; the later check has error 6.02 mV and fails the two-mV tolerance.

Observed 145.0 mV; error +6.02 mV.

A drift does not identify its cause

An indicator electrode and reference electrode contact the same sample and connect to a high-input-impedance meter. There is negligible net measurement current.

Conditioning, junction, temperature and matrix changes are candidates.

Uncertainty changes shape through the inverse

Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.

Symmetric voltage limits can become asymmetric concentration limits.

Place other electrochemical signals

Potential: potentiometry. Current: amperometry or voltammetry. Charge: coulometry.

Transfer: a 17.8 mV decrease

Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.

Predict the concentration factor before calculating.

Would this measure total fluorine?

Activity is dimensionless and related to concentration by an activity coefficient and standard concentration. Matching ionic strength supports calibration transfer.

Free-ion response does not automatically include every chemical form.

Take the reasoning into R

By hand, embedded experiment and R tasks · Practice quiz.

Save the calculation and a chemical interpretation in your reproducible report.

Interactive experiment · Predict, change, explain.