Electrodes and potentiometry
Activity, calibration and the complete cell
About this lecture
- Explain why the complete cell and activity conditions matter.
- Fit and invert a logarithmic fluoride calibration in R.
- Separate a shared drift failure from sample-specific matrix mismatch.
A voltage belongs to a cell
![An indicator electrode and reference electrode contact the same sample and connect to a high-input-impedance meter. There is negligible net measurement current.]()
Which two functions are required?
The indicator responds; the reference anchors
![An indicator electrode and reference electrode contact the same sample and connect to a high-input-impedance meter. There is negligible net measurement current.]()
The measured potential also includes the reference and junction.
Activity gives the thermodynamic response
![Activity is dimensionless and related to concentration by an activity coefficient and standard concentration. Matching ionic strength supports calibration transfer.]()
a_i=\gamma_i c_i/c^\circ.
Write the logarithmic relationship
E=K+(2.303RT/zF)\log_{10} a_i.
Anion and cation slopes need a convention
![Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.]()
For this fluoride example, increasing concentration lowers the potential.
Match conditions before using concentration
![Activity is dimensionless and related to concentration by an activity coefficient and standard concentration. Matching ionic strength supports calibration transfer.]()
Ionic-strength adjustment helps; it does not remove every interference.
pH is already logarithmic
![Illustrative pH potential decreases by 59.16 mV per pH unit at 25 degrees Celsius; the cell sets its intercept.]()
pH=-\log_{10}a_{H^+}.
Buffers establish the pH calibration
![Illustrative pH potential decreases by 59.16 mV per pH unit at 25 degrees Celsius; the cell sets its intercept.]()
Use appropriate buffers and measurement temperature.
Inspect the fluoride standards
![The same fluoride measurements form a curved relationship on an untransformed concentration axis.]()
Range here: 10⁻⁵ to 10⁻³ mol/L at matched conditions.
The correct axes reveal the model
![Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.]()
The same observations are nearly linear against log concentration.
Read the slope per decade
![Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.]()
Fitted slope = −59.10 mV per decade.
Invert in two steps
First (E-a)/b; then c=10^{(E-a)/b} mol/L.
Report the accepted early sample
![Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.]()
121.2 mV → 2.00 × 10⁻⁴ mol/L.
Check the calibration at a known concentration
![The early fluoride check has error 0.12 mV and passes; the later check has error 6.02 mV and fails the two-mV tolerance.]()
The check is 1.00 × 10⁻⁴ mol/L; exercise tolerance ±2.0 mV.
The early block passes
![The early fluoride check has error 0.12 mV and passes; the later check has error 6.02 mV and fails the two-mV tolerance.]()
Observed 139.1 mV; error +0.12 mV.
The later block fails
![The early fluoride check has error 0.12 mV and passes; the later check has error 6.02 mV and fails the two-mV tolerance.]()
Observed 145.0 mV; error +6.02 mV.
A drift does not identify its cause
![An indicator electrode and reference electrode contact the same sample and connect to a high-input-impedance meter. There is negligible net measurement current.]()
Conditioning, junction, temperature and matrix changes are candidates.
Uncertainty changes shape through the inverse
![Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.]()
Symmetric voltage limits can become asymmetric concentration limits.
Place other electrochemical signals
Potential: potentiometry. Current: amperometry or voltammetry. Charge: coulometry.
Transfer: a 17.8 mV decrease
![Fluoride potential decreases linearly with log concentration, with slope minus 59.10 mV per decade.]()
Predict the concentration factor before calculating.
Would this measure total fluorine?
![Activity is dimensionless and related to concentration by an activity coefficient and standard concentration. Matching ionic strength supports calibration transfer.]()
Free-ion response does not automatically include every chemical form.
Take the reasoning into R
By hand, embedded experiment and R tasks · Practice quiz.
Save the calculation and a chemical interpretation in your reproducible report.
Interactive experiment · Predict, change, explain.