Bleed-through and compensation

Image analysis

About this lecture

A two-colour fluorescence image: nucleus labelled green, membrane labelled orange. The orange channel shows a faint ghost of the nucleus. Nothing orange is in the nucleus: the green dye’s light leaks into the orange channel.

Two grey images of a synthetic cell: channel 1 shows the bright nucleus; channel 2 shows the membrane ring and a faint copy of the nucleus, marked "ghost of the nucleus".

  • Why dyes bleed through: broad emission spectra seen through finite filter bands.
  • Each dye gives one column of a mixing matrix: a pixel is \mathbf y = M\mathbf a.
  • A single-stained control measures a column.
  • Compensation undoes the mixing with M^{-1}; under- and over-compensation, and why similar dyes are hard.

Slides titled The math give the formulas and proofs; the ideas are complete without them.

Builds on: Day 46: Colour.

Why a dye shows up in the wrong channel

  • A dye emits over a broad range of wavelengths, with a long tail towards the red.
  • A channel’s filter lets through one band. Channel 1 is meant for dye 1, channel 2 for dye 2.
  • Dye 1’s tail reaches into channel 2’s band: a pixel with only dye 1 still gives a signal in channel 2. That is bleed-through (or crosstalk, spillover).
  • A second source: one laser can excite both dyes. The algebra below is the same.

A channel adds up a band

The value a channel records is the dye’s spectrum times the filter’s transmission, added up over wavelength.

Bar chart of two emission spectra in five 25 nm bins. Dye 1: 0.2, 1, 0.6, 0.3, 0.1 at 500 to 600 nm. Dye 2: 0, 0.1, 0.5, 1, 0.6. Channel 1's band covers 500 and 525 nm, channel 2's band 575 and 600 nm.

Five bins of 25 nm, one unit of each dye. Channel 1 passes the 500 and 525 nm bins (filter 1, else 0), channel 2 the 575 and 600 nm bins.

  • Dye 1: channel 1 sees 0.2 + 1.0 = 1.2, channel 2 sees 0.3 + 0.1 = 0.4 (times 25 nm).
  • Dye 2: channel 1 sees 0.1, channel 2 sees 1.6.
  • So one unit of dye 1 lands at (1.2,\ 0.4) in channel space, dye 2 at (0.1,\ 1.6).

The math: the channel value as an integral

With emission spectrum s_k(\lambda) of dye k (per unit amount) and transmission f_c(\lambda) of channel c (filter times detector sensitivity):

M_{ck} = \int s_k(\lambda)\, f_c(\lambda)\, d\lambda \;\approx\; \sum_i s_k(\lambda_i)\, f_c(\lambda_i)\, \Delta\lambda .

  • One column per dye: \mathbf m_k = (M_{1k}, \dots, M_{Nk})^\top, the dye as the channels see it.
  • A linear map from a spectrum to N numbers, the same as the camera of the colour day (from a spectrum to three numbers).
  • Narrower f_c: smaller off-diagonal M_{ck}, but also a smaller M_{kk}: fewer photons for the dye’s own channel.

Try it: spectra and filters

Fluorescence adds: the mixing matrix

Light from different dyes adds in each channel. A pixel with amounts a_1 of dye 1 and a_2 of dye 2 records

\mathbf y = a_1 \mathbf m_1 + a_2 \mathbf m_2 = M \mathbf a, \qquad M = \begin{pmatrix} \mathbf m_1 & \mathbf m_2 \end{pmatrix}.

  • Rows of M = channels, columns = dyes (components). Column k is where one unit of dye k lands: “the columns of a matrix are where the axes land” (the images unit, day 1).
  • Scale each column by its own channel: the spillover matrix S = \begin{pmatrix} 1 & s_{12} \\ s_{21} & 1 \end{pmatrix}. From the five-bin example: s_{21} = 0.4/1.2 = 0.33, s_{12} = 0.1/1.6 = 0.06.
  • The whole image: Y = MA, channels × pixels = (channels × dyes)(dyes × pixels). One matrix for all pixels.

Channel 2 against channel 1. Arrow m1 = (1.2, 0.4) in green, arrow m2 = (0.1, 1.6) in orange; from the tip of m1, half of m2 leads to the pixel y = (1.25, 1.2).

Brightfield stains multiply

Absorbing stains (H&E, DAB) do not add light, they remove a fraction of it: two stains on top of each other multiply the transmitted light.

  • Take the logarithm first, the optical density -\log_{10}(I/I_0): then the stains add, and the same model \mathbf y = M\mathbf a holds.
  • The unit on stain separation (colour deconvolution) does exactly that; today stays with fluorescence.

A single-stained control measures a column

Image a sample with only dye 1. Every pixel is a_1 \mathbf m_1: in the scatter plot of channel 2 against channel 1 the pixels lie on a line through the origin with slope s_{21}.

  • Four background-subtracted control pixels: (100, 27), (200, 48), (300, 77), (400, 98). The slope through the origin: \sum xy / \sum x^2 = 74\,600 / 300\,000 = 0.249.
  • Subtract the offset (dark level, background) first: an offset is a fake component with a flat spectrum and tilts the estimate.
  • A dim control gives a poor estimate: controls should be at least as bright as the sample.
  • One control per dye; with K dyes, K controls give the K columns.

Scatter plot of channel 2 against channel 1 for the four control pixels, with the line through the origin of slope 0.249.

The math: the slope through the origin

The control pixels satisfy y_i \approx s\, x_i. Least squares: minimise E(s) = \sum_i (y_i - s x_i)^2.

\frac{dE}{ds} = -2 \sum_i x_i (y_i - s x_i) = 0 \quad\Longrightarrow\quad \hat s = \frac{\sum_i x_i y_i}{\sum_i x_i^2}.

  • No intercept: zero dye gives zero signal (after offset subtraction).
  • Noise in x (the dye’s own channel) biases a regression slope towards 0; bright controls keep that bias small.

Compensation: undo the matrix

If \mathbf y = S\mathbf a, then \mathbf a = S^{-1}\mathbf y. With s_{21} = 0.25, s_{12} = 0.02:

S = \begin{pmatrix} 1 & 0.02 \\ 0.25 & 1 \end{pmatrix}, \qquad S^{-1} = \begin{pmatrix} 1.005 & -0.020 \\ -0.251 & 1.005 \end{pmatrix}.

  • The inverse has negative off-diagonal entries: “subtract 25 % of channel 1 from channel 2”, and a little the other way.
  • Compensation is one 2 \times 2 matrix applied to every pixel: the ghost of the nucleus disappears from channel 2.
  • It does not remove noise: the compensated values carry the noise of both channels, mixed through S^{-1}.

Channel 2 of the synthetic cell before and after compensation with the true spillover (0.25, 0.05): the grey ghost of the nucleus is gone, the membrane and vesicles remain.

Synthetic cell of the demo below (true spillover 0.25 and 0.05).

Compensation as pseudocode

s21 = Σ x·y / Σ x²   over a dye-1 control (x = channel 1, y = channel 2, offset removed)
s12 = the same over a dye-2 control, channels swapped
S⁻¹ = 1 / (1 − s12·s21) · (1  −s12 ; −s21  1)
for every pixel:
    (a1, a2) = S⁻¹ · (y1 − offset, y2 − offset)
  • Two numbers from two controls, then one 2 \times 2 matrix for every pixel.
  • Check: a single-stained structure must come out flat, at the background level.

The math: the 2 \times 2 inverse

\begin{pmatrix} a & b \\ c & d \end{pmatrix}^{-1} = \frac{1}{ad - bc}\begin{pmatrix} d & -b \\ -c & a \end{pmatrix}.

For S: \det S = 1 - s_{12}s_{21} = 1 - 0.02 \cdot 0.25 = 0.995, so S^{-1} = \frac{1}{0.995}\begin{pmatrix} 1 & -0.02 \\ -0.25 & 1 \end{pmatrix}.

  • \det S is the area of the parallelogram spanned by the two columns. Well-separated dyes: a near-square, \det \approx 1, S^{-1} \approx I minus small corrections.
  • Similar dyes: nearly parallel columns, a thin parallelogram, \det \to 0 and entries of S^{-1} of size 1/\det.

Try it: compensate the image

Under, over, and the check without ground truth

  • Under-compensated (guess too small): the single-stained population still tilts up; its median sits above the background.
  • Over-compensated (guess too large): it tilts down, below the background, into negative values.
  • A dye cannot emit negative light. Pixels of a single-stained population clearly below the background mean the spillover was set too high. A population that rises or falls with its own dye means the spillover is wrong.
Three scatter plots of the nucleus pixels (dye 1 only) after compensation with s21 guessed as 0.15, 0.25 and 0.35: the population tilts up with median 86, lies flat at median 10, or tilts down with median −66.

In the demo (true s_{21} = 0.25; dashed: the median): guess 0.15 leaves a median of 86 in channel 2 (background 11), guess 0.35 gives -66.

The math: what a wrong guess leaves

A pixel with only dye 1: \mathbf y = a\,(1,\ s)^\top (take s_{12} = 0). Compensating with the guess g:

\hat a_2 = y_2 - g\, y_1 = (s - g)\, a .

  • The dye-1 population lies on a line with slope exactly s - g: up when g < s, flat when g = s, down when g > s.
  • With both off-diagonals guessed (g_{12}, g_{21}): \hat{\mathbf a} = S_g^{-1} S \mathbf a; to first order in the errors the slope is still s_{21} - g_{21}.

Similar dyes: compensation amplifies noise

Two dyes with nearly the same spectrum: S_2 = \begin{pmatrix} 1 & 0.9 \\ 0.9 & 1 \end{pmatrix}, \det S_2 = 0.19,

S_2^{-1} = \begin{pmatrix} 5.26 & -4.74 \\ -4.74 & 5.26 \end{pmatrix}.

The columns of S (0.25 and 0.02 spillover) span a nearly square parallelogram, det 0.995; the columns of S2 (0.9 and 0.9) are nearly parallel and span a thin sliver, det 0.19.

  • A 10-unit error in channel 2 becomes errors of 47 and 53 in the two dyes. With S of the previous slides it stays 10.
  • The average is still right; individual pixels are not. In the demo the background noise grows from 4.6 to 31 after compensation.
  • Choosing dyes and filters that are far apart matters more than any algorithm. The next unit measures this with the condition number.

Bleed-through fakes colocalisation

Nucleus and membrane are different places, yet in the measured image the two channels correlate: every bright nucleus pixel is also a bit bright in channel 2.

  • In the demo: Pearson’s r between the channels inside the cell is 0.33 before compensation and −0.03 after.
  • Any measurement that compares channels (colocalisation, ratios, FRET) must compensate first. Colocalisation itself is taught in the colocalisation day; FRET in the FRET day.
  • Physical remedies: narrower filters (cost: photons), sequential excitation (one laser at a time), dyes further apart.

Aside: the same thing in flow cytometry

Photo: Mark Fergus, CSIRO, CC BY 3.0, via Wikimedia Commons.

  • A flow cytometer measures thousands of cells per second, each a “pixel” with one value per detector.
  • Same spillover, same S^{-1}, one compensation per cell. This is where “compensation” got its name.
  • Spectral cytometers unmix 20 or more dyes with many detectors: the least squares of the next unit.

Summary

  • Broad emission spectra and finite filter bands: each dye shows up in several channels (bleed-through).
  • A channel value is spectrum × filter, added over wavelength; one dye gives one column of M; fluorescence adds: \mathbf y = M\mathbf a.
  • A single-stained control measures a column: the slope through the origin, after offset subtraction.
  • Compensation: \hat{\mathbf a} = M^{-1}\mathbf y. A single-stained population should lie flat at the background level; tilted up = under, tilted down and negative = over.
  • Similar dyes: \det M \to 0, the inverse is large and amplifies noise.
  • Bleed-through makes independent channels correlate: compensate before comparing channels.