Fast MRI with compressed sensing

Image analysis

About this lecture

Magnetic resonance imaging is the flagship application of compressed sensing: the scanner measures Fourier coefficients, measuring them takes time, and we choose which ones to measure.

  • What the scanner measures (k-space), and why scan time grows with the number of lines.
  • Three ways to skip lines: the centre (blur), every other line (fold-over), random lines denser at the centre (noise-like artefacts).
  • The reconstruction: data consistency on the measured lines, wavelet sparsity elsewhere.
  • What can go wrong: small faint details are removed; too much acceleration fails abruptly.

Slides titled The math give the formulas and proofs; the ideas are complete without them.

MRI measures Fourier coefficients

One concept slide, no physics:

  • Magnetic field gradients make the signal of each position oscillate at a frequency that depends on where it is.
  • The received signal at one moment is therefore one Fourier coefficient of the object.
  • The scanner fills k-space (the 2D spectrum) one line at a time; the image is the inverse 2D DFT (recall the Fourier unit).

Most of the energy sits near the centre of k-space (the slow variations); edges and fine detail live further out.

Left: the synthetic head phantom, an ellipse with a bright rim and darker inner structures. Right: the logarithm of the magnitude of its 2D Fourier transform, bright at the centre and fading outwards, with one horizontal line marked in red: one readout.

Time = lines

  • Within one line (one readout) the samples come fast. Each new line needs a new excitation: a fixed repetition time per line.
  • A 3D scan with 256 × 128 phase-encoding lines at 10 ms per line: 256 \cdot 128 \cdot 0.01\,\text{s} = 328\,\text{s} = 5.5 min.
  • Patients move, children cannot lie still, the heart beats, contrast agents wash out. Fewer lines = shorter scans.
  • 4× fewer lines: 82 s. But which lines may we skip?

Bar chart of scan time against acceleration for 256 × 128 lines at 10 ms each: 5.5 min at 1×, 2.7 min at 2×, 1.4 min at 4×, 55 s at 6×, 41 s at 8×.

Three ways to skip lines

Top: three k-space masks (white = measured lines). Centre only: a band of lines in the middle. Every second line: alternating lines plus a full centre. Random lines, denser at the centre. Bottom: the zero-filled images. Centre only: a blurred phantom with ringing. Every second line: the phantom plus a ghost copy shifted by half the field of view. Random: the phantom with noise-like streaks.

  1. Centre only: a low-pass filter. The image is blurred with ringing (recall the Fourier unit). No reconstruction can bring the edges back: they were never measured.
  2. Every second line: the image folds: a ghost copy shifted by half the field of view (aliasing, as in the 1D copies of the last unit).
  3. Random lines, denser at the centre: noise-like artefacts, which a sparsity prior can remove.

Only option 3 fits compressed sensing.

The math: k-space sampling and the PSF of a mask

Measurements: \mathbf y = PF\mathbf x + \mathbf n, with F the orthonormal 2D DFT and P the rows (k-space samples) kept.

  • Zero-filled image: F^{-1}P^\top\mathbf y = \mathbf x * \mathrm{psf} (noise aside), where \mathrm{psf} = F^{-1}(\text{mask}) (unit 3).
  • A mask that keeps every r-th line in k_y: in y the PSF is r equal spikes N/r apart, so the image is the average of r copies shifted by FOV/r (field of view). For r = 2: one ghost at half the FOV.
  • Centre only (|k_y| < c/2): the PSF is a sinc-like (Dirichlet) kernel of width about N/c along y: blur plus ringing.
  • Random lines: a central peak plus low, irregular sidelobes along y.

The reconstruction

Three images of the phantom: fully sampled; zero-filled from 25 % of k-space with variable-density random lines, streaky, 21.3 dB; the compressed-sensing reconstruction with Haar wavelet ℓ1 from the same data, clean edges with some blockiness, 27.1 dB.

FISTA with the partial-Fourier forward operator and Haar-wavelet ℓ1 from 25 % of the lines: 21.3 → 27.1 dB. The measured lines are kept consistent with the data; the prior fills the rest with the sparsest image in wavelets.

CS-MRI reconstruction as pseudocode

x = zero-filled image
repeat:
    K = 2D DFT of x
    on every measured line: K = the measured data      # data consistency
    x = inverse 2D DFT of K
    c = wavelet transform of x
    soft-threshold the detail coefficients of c at λ   # sparsity
    x = inverse wavelet transform of c
  • The first three lines are the ISTA gradient step with step 1; the last three are the threshold. FISTA adds momentum, nothing else.
  • Per iteration: two FFTs and a wavelet transform pair.

The math: the CS-MRI problem

\min_{\mathbf x}\ \tfrac12\|PF\mathbf x - \mathbf y\|^2 + λ_1\|W\mathbf x\|_1 + λ_2\,\mathrm{TV}(\mathbf x) (Lustig, Donoho & Pauly 2007), with W an orthonormal wavelet transform.

  • PF has orthonormal rows (PFF^{-1}P^\top = I), so \|PF\| = 1, L = 1 and the ISTA step τ = 1 is safe.
  • One FISTA step: one FFT and one inverse FFT (the gradient step \mathbf x + F^{-1}P^\top(\mathbf y - PF\mathbf x): replace the measured k-space lines by the data), then the proximal step: a wavelet transform pair with soft thresholding; with TV as well, a few TV-denoising iterations (applying the two proximal steps one after the other is an approximation).
  • Real scanners use several receiver coils: \mathbf y_c = PFS_c\mathbf x with coil sensitivities S_c; the same algorithm with a larger operator.

Try it: compressed-sensing MRI

The math: variable density

Lines are drawn at random with probability p(k) \propto \left(1 - \frac{|k|}{k_{\max}}\right)^q, plus a fully sampled centre (here N/16 lines).

  • Why denser at the centre: the energy is there (77 % of the phantom’s energy within the central 17 of 256 lines), and coarse wavelets are coherent with low frequencies: a coarse wavelet is concentrated at low |k|, so random sampling there would miss most of its information.
  • Fine wavelets are spread over high frequencies, where sparse random sampling is enough.
  • Larger q: more concentrated at the centre (sharper drop).

Relative probability of sampling a line against |k|/kmax for q = 1 (a straight line), q = 3 and q = 6 (steeper), with the fully sampled centre shaded.

In practice

  • Compressed-sensing acceleration of about 2–8× is used clinically (commercial implementations since about 2017), mostly combined with parallel imaging: several receiver coils, each seeing the object through its own sensitivity profile, which adds equations.
  • Radial and spiral trajectories: radial lines through the centre of k-space are projections of the object (the Fourier slice theorem of back-projection); undersampled radial data give streaks, which are also incoherent.
  • Dynamic MRI (the beating heart, contrast uptake): successive frames differ little, so the series is very sparse in time; accelerations of 10× and more.
  • The reconstruction takes seconds to minutes per scan: computation replaces scan time.

What can go wrong

The phantom with five small faint discs (contrast 2–5 %), shown with a narrow grey window. CS at 4×: the large structures are right (27.1 dB), but the five faint discs have disappeared into a blocky texture. CS at 8×: the whole image breaks into blocks, 21.0 dB.

  • Small, low-contrast lesions vanish, already at 4× here, although the PSNR is good: the ℓ1 shrinkage of the sparsity unit removes small coefficients, and a faint 3-pixel lesion has only small coefficients.
  • TV makes a cartoon; Haar makes blocks.
  • Too much acceleration fails abruptly (the phase transition): at 8× the reconstruction breaks down.
  • Unlike a learned prior, a sparsity prior does not invent plausible anatomy; but it can remove real detail without any visible sign.

Learned reconstruction

  • Unrolling: write a fixed number of ISTA iterations as layers, and learn the thresholds and filters (or replace the threshold by a small network) from pairs of undersampled and fully sampled scans: “learned ISTA” (Gregor & LeCun 2010; for MRI, variational networks). Recall convolutional networks.
  • Diffusion priors for the same inverse problem: sample images that are consistent with the measured k-space lines and look like the training data (diffusion priors).
  • Faster and often better than hand-made sparsity, but a learned prior can hallucinate plausible structures that were not in the patient; a hand-made one mostly removes detail. Both need validation against fully sampled scans.

Summary

  • The MRI scanner measures Fourier coefficients, line by line; scan time is proportional to the number of lines.
  • Skipping lines: centre only blurs, regular undersampling folds (ghosts at FOV/r), random, variable-density undersampling gives noise-like artefacts.
  • CS-MRI: minimise misfit on the measured lines + λ·(wavelet ℓ1, TV) with FISTA; \|PF\| = 1, so the step is 1; each iteration costs two FFTs and a wavelet transform pair.
  • Variable density: sample the low frequencies densely (energy, coherence of coarse wavelets).
  • Clinically 2–8×, combined with multiple coils. The risk: small faint details are removed, and too much acceleration fails abruptly.
  • Next: instead of a fixed wavelet basis, learn the atoms from the images themselves.