Move two equal points closer together: the dip between their peaks shrinks, then vanishes.
Two Airy patterns (grey) and their sum (blue), NA 1.4, λ 520 nm.
Rayleigh: the peak of one point on the first dark ring of the other, d = 0.61\lambda/\mathrm{NA}. The dip is 26 %.
Sparrow: the dip just vanishes, d = 0.47\lambda/\mathrm{NA} (176 nm at NA 1.4).
Resolution is a convention. In practice noise decides: a 26 % dip on top of 2 photons per pixel is invisible.
The math: Rayleigh and Sparrow
Two equal points at \pm d/2, profile through both: I(x) = h(x - d/2) + h(x + d/2).
Rayleigh: h is the Airy pattern [2J_1(v)/v]^2 with v = 2\pi\mathrm{NA}\,r/\lambda; put d on its first zero, v = 3.832. At the midpoint each point contributes h at v = 1.916: [2J_1(1.916)/1.916]^2 = 0.367, so I(0) = 0.735 against a peak of about 1: a dip of 1 - 0.735 = 26\,\%.
Sparrow: the midpoint changes from a maximum to a minimum when its curvature changes sign: I''(0) = 2h''(d/2) = 0. The two profiles add up flat at the midpoint exactly when d/2 is at the inflection point of h.
Gaussian PSF h \propto e^{-x^2/2\sigma^2}: h'' = 0 at x = \sigma, so Sparrow is d = 2\sigma. With \sigma = 0.21\lambda/\mathrm{NA}: 0.42\lambda/\mathrm{NA}. For the Airy pattern, solved numerically: 0.47\lambda/\mathrm{NA}.
Try it: two points on pixels
Recall: two samples per cycle
From the frequency unit: a signal must be sampled at least twice per period of its finest component; faster patterns alias to false slow ones. That slide already said: the optics of a microscope should match the pixel size. Now we can say how.
The optics deliver nothing above2\mathrm{NA}/\lambda: the image is band-limited.
Two samples per period of the finest component: pixel \Delta x \le \dfrac{1}{2}\cdot\dfrac{\lambda}{2\mathrm{NA}} = \dfrac{\lambda}{4\mathrm{NA}}. Nyquist for a microscope.
The pixel that counts is the pixel in the sample plane: camera pixel ÷ magnification.
The finest pattern an NA 1.4 objective passes at 520 nm (period λ/2NA = 186 nm, grey). Sampled every 93 nm, it is recorded faithfully; sampled every 140 nm, the samples fit a false pattern with period 569 nm (dashed).
Nyquist in numbers
Camera with 6.5 µm pixels, λ = 520 nm:
objective
pixel in the sample
Nyquist \lambda/4\mathrm{NA}
verdict
100× / 1.4
65 nm
93 nm
fine (3.5 pixels across the Rayleigh distance)
60× / 1.4
108 nm
93 nm
slightly undersampled (1.2×)
20× / 0.75
325 nm
173 nm
undersampled by 1.9×
10× / 0.3
650 nm
433 nm
undersampled by 1.5×
Undersampled: fine periodic structure aliases (false stripes, moiré), and the camera, not the lens, sets the resolution.
Low-magnification objectives are often undersampled on purpose: a large field of view matters more there.
Two points at the Rayleigh distance (227 nm, NA 1.4) recorded with 65, 108 and 325 nm pixels (the first three rows of the table); a pixel sits on the midpoint, no noise.
The math: Nyquist from the OTF support
The image is g = f \ast h, so G = FH and G(\mathbf k) = 0 wherever H(\mathbf k) = 0: the image is band-limited to |\mathbf k| \le k_c = 2\mathrm{NA}/\lambda.
Sampling theorem: a band-limited function is determined by samples with spacing \Delta x \le 1/(2k_c) = \lambda/(4\mathrm{NA}) (per axis).
Along the optical axis, the widefield OTF extends to k_z = \dfrac{n}{\lambda}(1 - \cos\alpha), with \sin\alpha = \mathrm{NA}/n. So \Delta z \le \frac{\lambda}{2n(1 - \cos\alpha)}. NA 1.4 in oil (n = 1.518): \cos\alpha = 0.386, \Delta z \le 279 nm. NA 0.8 in air: 650 nm.
For small NA, 1 - \cos\alpha \approx \mathrm{NA}^2/2n^2, so \Delta z \le n\lambda/\mathrm{NA}^2.
Oversampling is not free
Smaller pixels than Nyquist add no detail (the optics deliver none), but they cost:
The same photons are spread over more pixels: fewer photons per pixel, a noisier pixel (shot noise \sqrt N, plus read noise per pixel; recall the camera noise unit).
Larger files, slower acquisition, and more bleaching per useful piece of information.
Example (demo): two points 300 nm apart, 500 photons each. With 65 nm pixels the peak pixel holds about 50 photons (noise ±14 %); with 20 nm pixels about 6.5 photons (±39 %).
The usual compromise: 2–3 pixels across the Rayleigh distance. Deconvolution (next day) needs at least Nyquist.
The same two points (300 nm apart, 500 photons each, Poisson noise) on 65 nm and 20 nm pixels. Bottom: the row through both points; blue the expected counts, orange one noisy recording.
The PSF in 3D
Left: widefield PSF in an xz slice (log scale), NA 1.4. Middle: confocal PSF. Right: the widefield OTF in the kx–kz plane: no information in a cone around the kz axis.
The PSF is elongated along the optical axis z. NA 1.4: FWHM 191 nm across, 515 nm along (2.7×); NA 0.8 in air: 335 nm and 1.17 µm (3.5×).
Rule of thumb for the axial extent: 2n\lambda/\mathrm{NA}^2 (for small NA; at NA 1.4 the exact value, 0.56 µm, is smaller than the rule’s 0.8 µm).
Widefield: every plane carries the same total light; out-of-focus light is spread, not removed. The OTF has a missing cone: no optical sectioning.
Confocal: a pinhole rejects out-of-focus light; PSF ≈ (excitation PSF) × (detection PSF): narrower, and the cone is filled.
Try it: the axial PSF
Measuring a PSF from beads
Image sub-resolution fluorescent beads (100 nm or smaller), then:
subtract the background (camera offset, autofluorescence);
find each bead’s centre with sub-pixel accuracy (e.g. its centroid);
shift every bead to a common centre and average many of them;
normalise to sum 1. Report the FWHM.
What limits it:
Bead size: the image is PSF ∗ bead. A bead of diameter D widens the measured PSF roughly like an extra blur of FWHM 0.53D: \text{FWHM}_\text{meas}^2 \approx \text{FWHM}_\text{PSF}^2 + (0.53D)^2.
Noise: averaging N beads reduces it by \sqrt N. Background left in widens the PSF. Pixels coarser than Nyquist blur the average.
Measured vs computed: the measured PSF contains the real aberrations of your microscope (often asymmetric along z); a computed one is noise-free but idealised.
Simulated bead slide: 30 beads of 40 nm, NA 1.4, 65 nm pixels, 500 photons per bead, background 10. One bead is noisy; the average of all 30 (aligned on a quarter-pixel grid) is smooth.
Averaging beads as pseudocode
for every bead:
subtract the background
(cx, cy) = centroid of the pixels around its brightest pixel
for every point (u, v) of a fine grid around 0:
sum(u, v) = sum(u, v) + I(cx + u, cy + v) # between pixels: interpolate
PSF = sum / (total of sum)
FWHM = width of the PSF's profile where it crosses half its peak
The fine grid (e.g. a quarter pixel) works because every bead sits at a different sub-pixel position.
Forgetting the first line widens the PSF: the background is averaged and normalised with it.
The math: averaging beads
Bead i gives b_i = h + e_i after centring, with independent noise of variance \sigma^2 per pixel. The mean \bar b = h + \tfrac1N\sum e_i has noise variance \sigma^2/N: standard deviation \sigma/\sqrt N. 30 beads: 1/\sqrt{30} = 0.18 of one bead’s noise.
A bead is a uniform sphere of radius R; seen from above, its brightness is the projected thickness \propto \sqrt{R^2 - r^2}. The image is PSF ∗ projection.
Variances add under convolution (for any normalised, centred profiles). Along x, a uniform sphere has variance R^2/5. A Gaussian with that variance has FWHM 2.355\,R/\sqrt5 = 0.53D. Treating the PSF as Gaussian too: \text{FWHM}_\text{meas}^2 \approx \text{FWHM}_\text{PSF}^2 + (0.53\,D)^2.
A 100 nm bead and a 191 nm PSF: \sqrt{191^2 + 53^2} = 198 nm (+4 %). A 200 nm bead: 218 nm (+14 %).
Try it: beads
Summary
Rayleigh0.61\lambda/\mathrm{NA} (26 % dip), Sparrow0.47\lambda/\mathrm{NA} (no dip): conventions. With few photons, noise decides.
Nyquist for a microscope: pixel in the sample plane \le \lambda/4\mathrm{NA} (93 nm at NA 1.4, 520 nm). Undersampling aliases; oversampling costs photons per pixel.
3D: the PSF is about 3–6× longer along z than across; axial Nyquist \lambda/2n(1 - \cos\alpha). Widefield has a missing cone; confocal removes out-of-focus light.
Beads: background subtraction, sub-pixel centring, averaging (\sqrt N), normalisation. Bead size adds about (0.53D)^2 to the squared FWHM.