Olink Explore — proximity extension assay with NGS readout
Protocol relationships
Some predecessor or candidate links are withheld until the corresponding protocol pages are available.
No formal ancestor assigned in this review. Plausible earlier chemistry analogue; descent or inspiration cannot be substantiated. Proximity-dependent antibody-DNA signal conversion followed by NGS readout; OlinkExplore uses PEA extension instead of the earlier PLA joining reaction.
Matched antibody–oligo pairs co-bind a protein, hybridize in proximity and template polymerase extension. Two PCR stages retain the paired assay barcodes and add an inline sample index for counting.
Olink publishes the 148-bp element order and run layout. Complete probe, assay-barcode, hybridization-arm, sample-index and PCR-primer sequences remain unverified in the source audit. Only the Illumina end is rendered base by base: the 20-nt P5 and 24-nt P7 portions are printed in Olink's patent, while the linker and Rd1SP bases following P5 are INFERRED as the standard Illumina TruSeq P5 end (consistent with the published 148-bp length and 66-cycle read, but not printed by Olink). The other regions remain structured roles.
Wording of the common forward-primer identity differs.
Wik et al. 2021, Experimental Procedures, PCR Amplification (PMC8633680 JATS line 168): forward PCR1 primer "complementary to Illumina's P5 adapter sequence"
Olink patent WO2021191442A1, worked PCR1/PCR2 example (Google Patents text lines 4135–4171): forward common primer "comprises the Illumina P5 sequencing adapter sequence (SEQ ID NO: 1)"
Modelled: The model gives the forward primer the P5 sense sequence, as the patent states and as the article's probe design requires (the forward probe itself carries P5 and Rd1SP). Evidence: Both statements come from Olink originals; the article also calls the reverse primer "complementary" to a common probe sequence, suggesting loose wording. Remaining uncertainty: Neither source prints the complete forward-primer bases, so its full sequence and any 5' extension beyond the 20-nt P5 portion remain unverified.
Reaction workflow
bead-purified 148-bp Olink Explore library
Bind matched oligo-antibody probes to the same protein. Inputs: matched antibody probe pair; conjugated forward and reverse PEA oligos; 0.6 µl block incubation mix + 0.2 µl sample, 4 °C for 16–24 h. Other outputs: free or non-proximal probes left in the well (no wash described) → unresolved
Extend across the hybridized probe arms. Inputs: polymerase in the Olink PCR1 mix (initial extension 50 °C, 20 min)
Preamplify every extended reporter with universal primers. Inputs: universal PCR1 forward primer; universal PCR1 reverse primer; Olink PCR1 enzyme
Pool PCR1 products from abundance blocks A–D. Inputs: equal-volume abundance-block pooling rule
Add the inline sample index and P7 end in ten PCR cycles. Inputs: PCR2 forward primer (same species as PCR1); sample-indexed P7 reverse primer; Olink PCR2 enzyme
Pool every PCR2 sample of a panel into one of four libraries. Inputs: pool all PCR2 samples of one panel (epMotion)
Purify each panel library with AMPure XP beads. Inputs: AMPure XP beads, 80 µl per 50 µl library. Other outputs: primers, primer dimers, dNTPs and other unbound contaminants → discard
Other material outcomes
free or non-proximal probes left in the well (no wash described) — unresolved
No removal is described: the PCR1/extension mix is added to each incubation reaction with no intervening wash, so these probes are inferred to be carried into the extension/PCR1 reaction. Wik et al. state only that proximal matched pairs generate detectable amplicons; the model does not follow these oligos further or assign their removal to the later AMPure XP cleanup.
Molecular structure is not represented in this graph.
primers, primer dimers, dNTPs and other unbound contaminants — discard
Removed in the bead supernatant before sequencing.
Molecular structure is not represented in this graph.
Proximity-extension topology
Two antibody–oligo conjugates must bind the same analyte closely enough for their terminal arms to pair. Polymerase then copies across the paired arms, joining the two assay identities in one reporter. Probe bases are proprietary and are therefore shown by role only.
Final sequencing library
INFERRED — linker and Rd1SP bases taken from Illumina's standard P5 end, not printed by Olink; the 66-nt inline length follows from that inferred 58-nt end. Published 148-bp element order: P5, the Read 1 site, a proprietary inline region containing both assay barcodes, the copied probe junction and sample index, then P7.
Sequencing primers
A Read 1 primer binds the published Rd1SP site (bases inferred as Illumina's standard Read 1 site). The NovaSeq run acquires one 66-cycle read and no separate index read; assay and sample identities are inline.
Sequences come from lib/; the landing site and the first bases of each read are computed from this page's final library, so they cannot drift from it.