Solid: source-documented · dashed: inferred · dotted gray: guessed earlier chemistry analogue. Guesses are not established ancestry; no link establishes identical chemistry.
A bead-immobilized ligand-binding oligo (LBO) holds a barcoded short release oligo (SRO) by a short stem; cognate ligand switches the LBO into a stem-loop and releases the SRO. Released and boiled-off SROs are amplified in two PCR rounds and counted by sequencing.
The drawn sensor is the mefloquine bSSA with barcode 1; nine barcode SROs (ATP, mefloquine, piperaquine; three barcodes each) were pooled in the sequenced experiment. The SRO's random N5–7 diversity region is drawn as 6 N. INFERRED — i5/i7 index lengths and the sequencing primers are not stated (the paper reports ~4 million 50-bp single-end MiSeq reads); the standard TruSeq Read 1 primer is inferred from the source-built Read 1 site and index windows are drawn as 8 nt.
Key oligos
Supplementary Data 1, mefloquine LBO (5′→3′): /5Biosg/ATACCGCTCTCGGGACGACGGCAGTCTATACCCCGTATCGCCGAAGGTTGTCGTCCC
Round 1: Reverse P5 primer ACACTCTTTCCCTACACGACGCTCTTCCGATC; Forward MQ primer GTGACTGGAGTTCAGACGTGTGCTCTTCCGATCT CGTCGTCCCGAGAGTAGC
Round 2: Forward i5 primer AATGATACGGCGACCACCGAGATCTACAC [i5 index] ACACTCTTTCCCTACACGACG; Reverse i7 primer CAAGCAGAAGACGGCATACGAGAT [i7 index] GTGACTGGAGTTCAGACGTG
Reaction workflow
released sequencing library
Anneal LBO and SRO (95°C 5 min, slow cool to 25°C). Inputs: mefloquine barcode-1 SRO (125 pmol)
Bind to streptavidin beads 30 min, wash twice, 45 min, wash once. Inputs: Dynabeads MyOne Streptavidin C1 (5 µL per sample). Other outputs: unbound excess SRO → discard
Ligand binding (45 min, rotator) folds the LBO stem-loop and displaces the SRO; collect supernatant. Inputs: sample: 125 µM ATP, 30 µM MQ, 25 µM PQ or PBS buffer. Other outputs: beads with SROs still paired to unswitched LBOs
Pool replicates, add spike-in controls and purify the released single-stranded SROs. Inputs: Monarch PCR & DNA cleanup kit (oligonucleotide protocol). Other outputs: column flow-through and washes → discard
Clean up round-1 PCR and dilute 1:5. Inputs: Monarch PCR & DNA cleanup kit (PCR protocol); 1:5 dilution. Other outputs: primers and flow-through → discard
Round-2 PCR adds i5/i7 indices and complete P5/P7 (100 µL, 8 cycles). Inputs: Round-2 Forward i5 primer (index value selected per library); Round-2 Reverse i7 primer (index value selected per library); PCR polymerase (not named in Methods)
Clean up and size-select (NucleoMag 1.2× bead:sample). Inputs: Monarch PCR cleanup then NucleoMag NGS beads at 1.2×. Other outputs: small fragments in the bead supernatant → discard
boiled sequencing library
Anneal LBO and SRO (95°C 5 min, slow cool to 25°C). Inputs: mefloquine barcode-1 SRO (125 pmol)
Bind to streptavidin beads 30 min, wash twice, 45 min, wash once. Inputs: Dynabeads MyOne Streptavidin C1 (5 µL per sample). Other outputs: unbound excess SRO → discard
Ligand binding (45 min, rotator) folds the LBO stem-loop and displaces the SRO; collect supernatant. Inputs: sample: 125 µM ATP, 30 µM MQ, 25 µM PQ or PBS buffer. Other outputs: 'released' sample: SROs displaced into the supernatant
Boil off the remaining SROs to measure total sensor (percentage release). Inputs: Mg-free strand separation buffer, 95°C. Other outputs: beads carrying biotin-LBO → discard
Pool replicates, add spike-in controls and purify the boiled single-stranded SROs. Inputs: Monarch PCR & DNA cleanup kit (oligonucleotide protocol). Other outputs: column flow-through and washes → discard
Clean up round-1 PCR and dilute 1:5. Inputs: Monarch PCR & DNA cleanup kit (PCR protocol); 1:5 dilution. Other outputs: primers and flow-through → discard
Round-2 PCR adds i5/i7 indices and complete P5/P7 (100 µL, 8 cycles). Inputs: Round-2 Forward i5 primer (index value selected per library); Round-2 Reverse i7 primer (index value selected per library); PCR polymerase (not named in Methods)
Clean up and size-select (NucleoMag 1.2× bead:sample). Inputs: Monarch PCR cleanup then NucleoMag NGS beads at 1.2×. Other outputs: small fragments in the bead supernatant → discard
Other material outcomes
unbound excess SRO — discard
removed with the bead washes
Molecular structure is not represented in this graph.
beads carrying biotin-LBO — discard
magnetically retained; the supernatant is collected
Molecular structure is not represented in this graph.
column flow-through and washes — discard
column flow-through and washes
Molecular structure is not represented in this graph.
primers and flow-through — discard
column flow-through and washes
Molecular structure is not represented in this graph.
small fragments in the bead supernatant — discard
bead supernatant and washes
Molecular structure is not represented in this graph.
column flow-through and washes — discard
column flow-through and washes
Molecular structure is not represented in this graph.
primers and flow-through — discard
column flow-through and washes
Molecular structure is not represented in this graph.
small fragments in the bead supernatant — discard
bead supernatant and washes
Molecular structure is not represented in this graph.
Final sequencing library
INFERRED — 8-nt i5/i7 windows drawn; index lengths are not stated. MQ barcode-1 amplicon: Read 1 crosses the N5–7 diversity bases, then the 8-nt analyte barcode and the MQ stem.
Sequencing primers
MiSeq, ~4 million 50-bp single-end reads (Methods). Read 1 primer identity and index-read cycles are not stated; the TruSeq Read 1 primer is inferred from the source-built site.
Sequences come from lib/; the landing site and the first bases of each read are computed from this page's final library, so they cannot drift from it.
anneals to the bottom strand over round-1 reverse P5 primer / Read 1 site
NNNNNNGTCAGA… (from N5-7 complement)
Read content and handling
Content and cycle positions are derived from the final construct, primer direction, molecular-feature annotations and declared run lengths. ‘No’ means the read is auxiliary or absent, not that its information is unimportant.
Read
Cycles
Content
Include in sequence analysis?
Read 1
50