Atomic measurements
From elemental light to original dry mass
About this lecture
- Distinguish atomic absorption from optical emission.
- Inspect spectral overlap and independent quality checks.
- Undo dilution and correct the method blank on a common basis before reporting dry mass.
Absorption supplies the photon
![Absorption moves an atom from lower to higher energy, while emission releases a photon during a transition from higher to lower energy.]()
A source photon promotes a suitable free atom to a higher state.
Emission releases the photon
![Absorption moves an atom from lower to higher energy, while emission releases a photon during a transition from higher to lower energy.]()
An excited population emits characteristic light during transitions.
Follow the AAS path
![The sample must be converted into an appropriate free-atom population; flame and graphite furnace differ in how this is achieved.]()
Element source → atomiser → selected wavelength → detector.
The flame introduces material continuously
![Flame AAS, graphite furnace AAS and ICP optical emission differ in introduction, atomisation and detection conditions.]()
Transport and atomisation affect sensitivity.
The graphite furnace treats a small aliquot
![Flame AAS, graphite furnace AAS and ICP optical emission differ in introduction, atomisation and detection conditions.]()
A heating program separates drying, matrix removal and atomisation.
Follow the ICP optical-emission path
![Elemental emission does not preserve the original molecular identity or oxidation-state distribution of the sample.]()
Aerosol → plasma excitation → wavelength separation → detector.
Many elements means many possible lines
![Elemental emission does not preserve the original molecular identity or oxidation-state distribution of the sample.]()
Line choice and background still matter.
Look around the target line
![An illustrative neighbouring emission feature changes the spectrum around a target line; profiles are not identification references.]()
Which signal would a broad integration window include?
An interference need not look like a separate peak
![An illustrative neighbouring emission feature changes the spectrum around a target line; profiles are not identification references.]()
Transport and chemical effects can change the target population.
Calibrate the introduced solution
![Copper emission counts increase approximately linearly over zero to two mg/L in introduced solution.]()
Concentration axis: mg copper/L measured solution.
Check with independent material
![Two independent dissolved copper checks recover about 100.4 and 100.8 percent, within the exercise window.]()
Both supplied checks meet the exercise 95–105% rule.
Trace the sample mass and volumes
![A1: (0.40189 × 2 − 0.001992) mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry mass.]()
A1: 0.2500 g dry material → 50.00 mL digest → twofold dilution.
Convert the sample signal
![Copper emission counts increase approximately linearly over zero to two mg/L in introduced solution.]()
A1 measured solution ≈ 0.40189 mg/L copper.
The method blank has a different dilution history
![A blank, independent reference check and a sample spike examine different parts of the measurement.]()
Blank digest is measured undiluted.
Subtract on a common basis
![A1: (0.40189 × 2 − 0.001992) mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry mass.]()
A1 digest: 0.40189 × 2 − 0.001992 = 0.80179 mg/L.
Recover amount, then divide by original mass
![Conditional copper dry-mass results are about 160 and 300 mg/kg in two supplied plant samples.]()
0.80179 mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry.
A blank has no sample mass denominator
![Conditional copper dry-mass results are about 160 and 300 mg/kg in two supplied plant samples.]()
No mg/kg plant result is assigned to the method blank.
Passing a dissolved check cannot prove digestion recovery
![Two independent dissolved copper checks recover about 100.4 and 100.8 percent, within the exercise window.]()
What material would better challenge the whole preparation?
Uncertainty needs more than counting noise
![Averaging repeated readings reduces independent reading noise; it does not remove a common stock or preparation effect.]()
Mass, volume, calibration, blank variation and recovery can all contribute.
Choose among instruments for the question
![Flame AAS, graphite furnace AAS and ICP optical emission differ in introduction, atomisation and detection conditions.]()
Range, matrix, elements, sample volume, throughput and required uncertainty.
Transfer the blank and dry-mass basis
![A1: (0.40189 × 2 − 0.001992) mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry mass.]()
0.2000 g to 25.00 mL; fivefold dilution reads 0.0800 mg/L; blank digest 0.0100 mg/L.
Take the reasoning into R
By hand, embedded experiment and R tasks · Practice quiz.
Save the calculation and a chemical interpretation in your reproducible report.
Interactive experiment · Predict, change, explain.