Atomic measurements

From elemental light to original dry mass

Analytical Chemistry

About this lecture

  • Distinguish atomic absorption from optical emission.
  • Inspect spectral overlap and independent quality checks.
  • Undo dilution and correct the method blank on a common basis before reporting dry mass.

Which information survives atomisation?

Absorption moves an atom from lower to higher energy, while emission releases a photon during a transition from higher to lower energy.

An element amount can survive; original molecular identity generally does not.

Absorption supplies the photon

Absorption moves an atom from lower to higher energy, while emission releases a photon during a transition from higher to lower energy.

A source photon promotes a suitable free atom to a higher state.

Emission releases the photon

Absorption moves an atom from lower to higher energy, while emission releases a photon during a transition from higher to lower energy.

An excited population emits characteristic light during transitions.

Follow the AAS path

The sample must be converted into an appropriate free-atom population; flame and graphite furnace differ in how this is achieved.

Element source → atomiser → selected wavelength → detector.

The flame introduces material continuously

Flame AAS, graphite furnace AAS and ICP optical emission differ in introduction, atomisation and detection conditions.

Transport and atomisation affect sensitivity.

The graphite furnace treats a small aliquot

Flame AAS, graphite furnace AAS and ICP optical emission differ in introduction, atomisation and detection conditions.

A heating program separates drying, matrix removal and atomisation.

Follow the ICP optical-emission path

Elemental emission does not preserve the original molecular identity or oxidation-state distribution of the sample.

Aerosol → plasma excitation → wavelength separation → detector.

Many elements means many possible lines

Elemental emission does not preserve the original molecular identity or oxidation-state distribution of the sample.

Line choice and background still matter.

Look around the target line

An illustrative neighbouring emission feature changes the spectrum around a target line; profiles are not identification references.

Which signal would a broad integration window include?

An interference need not look like a separate peak

An illustrative neighbouring emission feature changes the spectrum around a target line; profiles are not identification references.

Transport and chemical effects can change the target population.

Calibrate the introduced solution

Copper emission counts increase approximately linearly over zero to two mg/L in introduced solution.

Concentration axis: mg copper/L measured solution.

Check with independent material

Two independent dissolved copper checks recover about 100.4 and 100.8 percent, within the exercise window.

Both supplied checks meet the exercise 95–105% rule.

Trace the sample mass and volumes

A1: (0.40189 × 2 − 0.001992) mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry mass.

A1: 0.2500 g dry material → 50.00 mL digest → twofold dilution.

Convert the sample signal

Copper emission counts increase approximately linearly over zero to two mg/L in introduced solution.

A1 measured solution ≈ 0.40189 mg/L copper.

The method blank has a different dilution history

A blank, independent reference check and a sample spike examine different parts of the measurement.

Blank digest is measured undiluted.

Subtract on a common basis

A1: (0.40189 × 2 − 0.001992) mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry mass.

A1 digest: 0.40189 × 2 − 0.001992 = 0.80179 mg/L.

Recover amount, then divide by original mass

Conditional copper dry-mass results are about 160 and 300 mg/kg in two supplied plant samples.

0.80179 mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry.

A blank has no sample mass denominator

Conditional copper dry-mass results are about 160 and 300 mg/kg in two supplied plant samples.

No mg/kg plant result is assigned to the method blank.

Passing a dissolved check cannot prove digestion recovery

Two independent dissolved copper checks recover about 100.4 and 100.8 percent, within the exercise window.

What material would better challenge the whole preparation?

Uncertainty needs more than counting noise

Averaging repeated readings reduces independent reading noise; it does not remove a common stock or preparation effect.

Mass, volume, calibration, blank variation and recovery can all contribute.

Choose among instruments for the question

Flame AAS, graphite furnace AAS and ICP optical emission differ in introduction, atomisation and detection conditions.

Range, matrix, elements, sample volume, throughput and required uncertainty.

Transfer the blank and dry-mass basis

A1: (0.40189 × 2 − 0.001992) mg/L × 0.05000 L / 0.0002500 kg = 160.36 mg/kg dry mass.

0.2000 g to 25.00 mL; fivefold dilution reads 0.0800 mg/L; blank digest 0.0100 mg/L.

Take the reasoning into R

By hand, embedded experiment and R tasks · Practice quiz.

Save the calculation and a chemical interpretation in your reproducible report.

Interactive experiment · Predict, change, explain.