Quiz: atomic measurements
Try each question before opening its hint. Then explain the answer in chemical and measurement terms. Lecture · Exercise.
Q1. Which optical signal distinguishes AAS from ICP-OES?
Ask whether the useful photons come from a source lamp or the excited sample.
AAS measures attenuation of suitable source light by free atoms. ICP-OES measures light emitted after plasma excitation and separates its wavelengths. ICP-OES is not ICP-MS, which detects ions by mass-to-charge.
Q2. A 0.5000 g dry sample is made to 25.00 mL digest. A fourfold dilution measures 0.300 mg/L copper; the undiluted method blank is 0.020 mg/L. Calculate mg Cu/kg dry.
Undo the sample dilution before subtracting the undiluted digest blank.
Digest is \(0.300\times4-0.020=1.180\) mg/L. Amount is \(1.180\times0.02500=0.02950\) mg. Divide by 0.0005000 kg to obtain 59.0 mg Cu/kg dry mass.
Q3. For the preceding sample, an analyst instead computes \((0.300-0.020)\times4\) as corrected digest concentration. Why is it wrong?
Was the method blank diluted fourfold?
The blank was measured undiluted. Subtracting it before multiplying by four subtracts four times its digest contribution rather than once. This gives 1.120 instead of 1.180 mg/L and biases the dry-mass result low by 3.0 mg/kg.
Q4. Counts at a selected wavelength repeat very consistently, but another element has a nearby line. Does the small SD establish copper selectivity?
Repeatability and attribution of a signal are different questions.
No. A reproducible overlapping contribution can give precise but biased copper estimates. Spectral/background evidence, appropriate alternative lines or justified corrections are needed; a method blank cannot automatically remove a sample-specific neighbour.
Q5. A dissolved copper check recovers 100%. Does it establish quantitative recovery from native plant material?
Which preparation steps did the dissolved check bypass?
No. It does not recreate release from the plant matrix or establish digestion completeness. A suitable reference material or relevant recovery evidence must challenge the native preparation, with its own assumptions.
Q6. Does a dry-mass copper concentration identify its original copper-containing molecule or oxidation state?
Atomisation changes the material.
No. The elemental result does not preserve that chemical identity. A specific species claim needs a species-preserving preparation and suitable selective evidence. The concentration basis must still distinguish digest mg/L from original dry mg/kg.