UV and visible absorption

Iron in water: from colour to concentration

Analytical Chemistry

About this lecture

A coloured solution gives an absorbance. We want mg elemental Fe/L in the supplied filtered water.

  • Explain what absorbs and how light is measured.
  • Fit a justified calibration and return to the original sample basis.
  • Decide what range, QC and uncertainty evidence permit us to report.

The chemical absorber

Iron is reduced to Fe(II), then complexed with excess 1,10-phenanthroline (phen).

We assume quantitative conversion and no analyte loss.

Iron(II) combines with three phenanthroline ligands to form a coloured complex.

What the detector measures

T=I/I_0, then A=-\log_{10}T.

A reagent blank establishes the reference under matched conditions.

Selected light passes through a coloured solution and transmitted intensity is measured.

Why select a wavelength?

A strong response near 510 nm supports useful sensitivity.

A maximum does not prove the sample has no other absorber.

An illustrative iron-standard spectrum has a maximum near 510 nm.

Beer–Lambert connects light and composition

A=\varepsilon l c at fixed wavelength, with appropriate chemical and instrumental conditions.

For a molar absorptivity, c is in mol/L. Our empirical line instead uses mg Fe/L.

Doubling absorbance does not halve transmission

T=0.80\Rightarrow A\approx0.09691.

If A doubles, T=10^{-A} becomes 0.64, not 0.40.

Selected light passes through a coloured solution and transmitted intensity is measured.

Fit the quantity we deliberately varied

Known final concentrations are the explanatory variable; absorbance is the response.

fit <- lm(absorbance ~ standard_mg_l, data = standards)
Prepared final-solution iron standards and their retained-intercept calibration.

Interpret the fitted relationship

A=0.003874+0.198799c_{final}.

The slope has units L/mg Fe; the offset is estimated from the data.

Prepared final-solution iron standards and their retained-intercept calibration.

Residuals ask whether the line misses structure

Residual = observed response − fitted response.

Inspect curvature, changing spread and unusual points before accepting the model.

Standard residuals plotted against prepared concentration.

Invert the response model

\hat c_{final}=(\overline A-a)/b.

P01 has \overline A=0.480620, giving 2.398 mg Fe/L final.

Count the experimental units

P01 and P02 are separate preparations of W1; each has three readings.

Six readings are not six independently collected waters.

One collected water sample gives two preparations, each with three readings.

Carry the preparation record into the calculation

c_{original}=c_{final} V_{final}/V_{aliquot}.

P01: 2.398\times50.00/10.00\approx 11.99 mg Fe/L original.

A ten millilitre aliquot is made up to fifty millilitres total.

A new dilution means a new measurement

P03 exceeds the calibrated response range.

P04 is a new 5.00 mL aliquot made to 50.00 mL and measured again.

An old out-of-range preparation and a newly measured dilution have different evidence.

Independent QC can block an in-range result

Check material: 2.00 mg Fe/L. Exercise acceptance: 95–105% recovery.

The later block B fails, so P05 is withheld.

Independent QC recovers 99.6 percent in block A and 85.2 percent in block B.

Uncertainty survives repeat readings

P01: 11.991 mg Fe/L, with standard uncertainty 0.028 mg Fe/L from the stated model.

That number is not a 95% half-width.

Contributions to P01 standard uncertainty from calibration and volumes.

A result needs its chemical assumptions

Which check addresses contamination? Which challenges the line? Which could detect a sample-specific matrix effect?

The chemical result connects question, preparation, calibration and evidence.

Explore the mechanism, then test a claim

Embedded experiment: change path or wavelength, then introduce stray light.

Predict the response before moving a control. Export a scenario and reproduce the calculation in R.

Standard residuals plotted against prepared concentration.

Take the complete argument into the exercise

By hand, demo and R tasks · Check your reasoning

Report the absorber, wavelength, concentration basis, preparation factor, checks and uncertainty limitations.

The chemical result connects question, preparation, calibration and evidence.