Quiz: UV and visible absorption

Answer before opening the hint or explanation.


Q1. What species absorbs, and what iron quantity is reported?

Separate the prepared complex from the original oxidation state.

The prepared Fe(II)–phenanthroline complex absorbs. With quantitative reduction and complex formation, the result describes elemental iron in supplied filtered water, not its original Fe(II) fraction.


Q2. Why does doubling absorbance not halve transmitted light?

Invert the logarithm.

\(T=10^{-A}\). Doubling absorbance squares transmittance. At \(A=0.30103\), \(T=0.50\); at \(A=0.60206\), \(T=0.25\).


Q3. Should already reagent-blank-referenced absorbances be blank-corrected again?

Read the measurement basis.

No. Another subtraction changes the stated basis. A whole-procedure method blank can investigate a different contamination question.


Q4. Why fit absorbance on prepared concentration?

Which variable is deliberately set?

Concentration is the controlled input and absorbance the response. Invert that fitted response relationship for the unknown; reverse regression is a different error model.


Q5. P04 uses 5.00 mL made up to 50.00 mL. What factor restores the original basis?

Final total volume divided by aliquot.

The factor is 10. The signal must belong to that newly measured preparation; editing a factor cannot rescue an old out-of-range reading.


Q6. A sample is in range but its block QC fails. Is it reportable?

Range and independent check answer different questions.

Withhold it, investigate the failure, establish satisfactory calibration and QC and remeasure. The QC recovery alone does not identify a valid correction.


Q7. P01 and P02 are two preparations from the same supplied water W1. Do three readings from each give six independently collected water samples?

Track the hierarchy.

No. They are repeated readings nested within two preparations of one supplied water. They share sample origin and calibration.


Q8. What does the stated uncertainty omit?

Ask what the mathematical model includes.

The example includes reading/calibration and supplied volume uncertainties under stated independence assumptions. Sampling, matrix bias, incomplete recovery and uncertain standard concentrations are excluded; it is not full method uncertainty.