Quiz: Calibration and unknown samples
Try each question before opening its hint or answer.
Q1. Known concentration is in mg/L; the measured response is absorbance. Which formula fits the calibration used here?
lm(absorbance ~ concentration_mg_l, data = standards)lm(concentration_mg_l ~ absorbance, data = standards)
The response is left of ~; the predictor is right of it.
(i). Prepared concentration is the predictor and absorbance the response. The unknown concentration is later estimated by rearranging that fitted line.
Q2. For absorbance = 0.04 + 0.065 * concentration_mg_l, estimate concentration from an in-range reading of 0.30. What does the slope mean?
Rearrange to (reading - intercept) / slope.
\((0.30-0.04)/0.065 = 4\) mg/L. The slope is an increase of 0.065 absorbance per mg/L increase in concentration.
Q3. A standard has observed absorbance 0.30 and fitted absorbance 0.28. What is its residual, and where is its point relative to the line?
Residual = observed − fitted.
The residual is +0.02 absorbance, so the point lies above the fitted line. It is a signal-scale difference, not a concentration.
Q4. Measured standard signals span 0.024–0.525. Unknown X1 reads 0.270 and X2 reads 0.700. Which gets a supported inverse estimate, and how should the other appear in the saved table?
Check the measured signal range before inverting.
X1 is in range. Retain X2’s ID and reading with an out-of-range flag and a missing concentration estimate. A suitable new measurement is needed.
Q5. A new sample is diluted from 1.00 mL to 10.00 mL total volume. Its new in-range reading corresponds to 7.40 mg/L. Estimate the original concentration assuming no measured substance is lost or reacts.
Multiply the diluted estimate by final total volume / original-sample volume.
\(7.40\times(10.00/1.00) = 74.0\) mg/L using the rounded estimate. This uses a measured dilution and stated volumes; it does not validate a different sample’s out-of-range reading.