Genomic DNA libraries

Use cases · NGS library preparation

Johan Henriksson

Start from the question, not the kit

A genomic library answers one of a few questions about a human genome:

  • Variants in a germline or tumour genome: needs even coverage and few artefacts
  • Copy number: needs coverage that tracks the template, not the chemistry
  • Ancient or forensic DNA: the molecules are short and damaged before you touch them
  • One cell: there are two copies of each locus, and you have to make more

This deck derives the library for each, then reads real protocols and asks why they differ.

Five blocks, and the sample picks the variant

  • The blocks are the ones from Part B: fragmentation, end chemistry, handle, amplification, QC
  • For every genomic protocol, ask which variant of each block it chose, and what in the sample forced it
  • Most differences between kits are a different answer to block 4: is there PCR?

Mass of DNA decides the route

Design: a clinical whole genome from 1 µg of blood DNA

Question: germline variants and copy number across the whole genome, GC-extremes included. Sample: a blood draw, so micrograms of long, intact DNA.

  • With that much input, block 4 can be nothing, and the other four blocks have to hand the sequencer a finished molecule on their own
  • The whole walk is above; the next four slides take it one block at a time, asking after each what sits on the two ends. The protocol is TruSeq DNA PCR-Free
  • NEBNext Ultra II FS PCR-free makes the same choices with enzymatic fragmentation; the same blocks with 5 to 12 cycles of PCR are the NEBNext Ultra II family

Block 1: shear the 1 µg to about 350 bp

  • Acoustic shearing is unbiased by sequence, which is what even coverage needs (fragmentation)
  • 350 bp is set against the read length and the cluster, not against the chemistry (what the sequencer wants)
  • Construct now: insert of about 350 bp, ragged ends, a mixture of overhangs and 5’-OH, which no ligase can use
  • Row 2 repairs and A-tails in the same tube; row 3 does blocks 1 to 3 at once, and returns later in this deck

Blocks 2 and 3: repair the ends, add one A, ligate the adapter

  • End repair leaves every end blunt with a 5’-phosphate, one 3’ A is added, and the adapter carries the matching T (end chemistry)
  • The A/T handshake is what keeps the product clean: inserts cannot join each other, and adapters cannot join each other
  • Construct now: adapter – insert – adapter, one at each end, the unpaired arms of each fork still single-stranded
  • Each of those enzyme steps ends in a bead clean-up, and every clean-up loses molecules

A-tailing and end repair as in Illumina’s TruSeq guide, see the collection page for TruSeq PCR-Free

Block 4: nothing, because the fork was already full length

  • A full-length adapter arrives with P5, i5 and R1 on one end and R2, i7 and P7 on the other, so the molecule is finished the moment it is ligated (the molecule)
  • Construct now, read out: P5-i5-R1-insert-R2-i7-P7. Every ligated fragment can seed a cluster, and no base of it has been copied
  • With stub adapters the flow-cell ends are missing and PCR installs them, so block 4 stops being optional
  • What that PCR costs — GC and length bias, duplicates, library complexity — is worked out in amplification

Kozarewa et al. 2009, doi:10.1038/nmeth.1311

Block 5: size select, count by qPCR, load

  • Bead size selection brings the trace to one peak at the design size; a sharp peak near 130 bp is adapter dimers, which would take cluster space
  • Fluorescence counts every molecule, including the ones that lost an adapter: qPCR counts only those carrying both flow-cell ends (cleanup and QC)
  • Construct at the flow cell: unchanged since block 3. The molecule that was ligated is the molecule that is read
  • The third trace belongs to the PCR route, where a second, larger peak means too many cycles

The same spine, blocks 1 to 3 in one enzyme: tagmentation

  • A Tn5 dimer loaded with adapters cuts the DNA and joins an adapter to each 5’ end (transposases)
  • Construct after that one step: adapter – insert – adapter, with a 9 nt gap on the other strand at each end
  • The adapters are stubs and the gap is unfilled, so the first PCR does both jobs: block 4 is not optional here
  • Less input and fewer steps, at the price of a sequence preference: Nextera XT and Illumina DNA Prep

Adey et al. 2010, doi:10.1186/gb-2010-11-12-r119

What the field does, side by side

  • Tagmentation kits give up PCR-free-ness for input and speed; ligation kits give up input for flatness
  • Illumina DNA Prep uses bead-linked transposomes, Tn5 held on beads so that a fixed number of enzyme sites does the cutting: yield saturates, and libraries pool without individual quantification above 100 ng
  • Nextera XT was designed around small genomes, and its 1 ng input is matched to them rather than to a human genome

Worked example: how many molecules do you really have?

10 ng of human DNA, 3.3 pg per haploid genome:

  • 10 000 pg / 3.3 pg is about 3000 copies of every locus
  • Assume 20% of fragments end up with an adapter at both ends (an assumption: check your protocol): about 600 molecules per locus
  • 30x sequencing: x = 30/600 = 0.05 reads per molecule, so about 97% unique on the saturation curve (amplification)
  • Now 100 pg: 30 copies, about 6 molecules per locus, x = 5 at 30x, and about 20% unique

At 100 pg, no extra sequencing rescues the library. Only more input or a different method does.

Degraded DNA is damaged before you start

  • Ancient and some forensic DNA, or DNA from FFPE blocks (formalin-fixed, paraffin-embedded tissue), is short, nicked, and full of deaminated cytosines (C becomes U, read as T)
  • The damage accumulates at molecule ends, so the read-end pattern is the signature that the DNA is authentic
  • Library prep must keep short molecules and decide what to do about the uracils

Briggs et al. 2007, doi:10.1073/pnas.0704665104

Deal with the uracils: UDG

  • Uracil-DNA glycosylase plus an endonuclease removes uracil, so calls become clean
  • Partial UDG removes it internally but leaves the ends, so authenticity can still be tested
  • See partial-UDG double-stranded ancient DNA

Rohland et al. 2015, doi:10.1098/rstb.2013.0624

Short and nicked: make the library from single strands

  • A ligation library needs both strands of a duplex intact at the ends (what a ligase requires); ancient DNA often is not
  • Denature first, ligate an adapter to every single strand, copy, then ligate the second adapter
  • Single-stranded ancient DNA library: more molecules recovered, at the cost of more steps

Gansauge and Meyer 2013, doi:10.1038/nprot.2013.038

One cell: there is nothing to prepare a library from

  • A cell holds about 6.5 pg: two copies of each locus
  • Amplification must produce nanograms to micrograms, and it can fail in the three ways shown
  • Every single-cell WGA method is a different attempt to make that amplification even and faithful

Amplifying one genome is not amplifying a library

  • A bulk library is already millions of distinct molecules and PCR makes more of each. Whole-genome amplification has to create the library first, roughly a million-fold from two copies, and only then do blocks 1 to 3 run
  • So unevenness that a bulk library would absorb arrives here as a genotype: a locus copied from one allele and not the other reads as a homozygote, and an error made in the first copy sits in half the reads
  • MDA, MALBAC and PTA are three attempts at that amplification, and they differ in how many generations stand between the original molecule and each copy (amplification); the product then enters an ordinary ligation library
  • What you can call follows from the evenness you get: copy number tolerates noise in depth, single-nucleotide variants tolerate neither dropout nor an early error

After sequencing: read the run back

  • Duplicate rate and insert-size distribution say whether the library held as many distinct molecules as the run asked for
  • The GC-versus-coverage curve says whether the chemistry tracked the template or its own preferences
  • Fold-80 base penalty: how much more depth you would need for 80% of positions to reach the mean coverage. 1 is perfectly even, 2 means half the run went on evenness
  • These are the numbers that answer the question on the first slide; the Bioanalyzer trace only said the library was built

Design from the sample, then compare

  • The recurring exercise: given the question and the sample, derive the library, then compare with the kit
  • Where your design and the kit differ, ask what the vendor knew about the sample that you did not
  • Next: targeted sequencing uses the same blocks and adds a selection step

Summary

  • The sample decides the route: micrograms of intact DNA allow a PCR-free ligation library, in which the ligated molecule is the sequenced molecule
  • Low input is what forces PCR or tagmentation, and amplification holds what that costs
  • Damaged DNA needs end and uracil choices; a single-stranded library recovers the shortest molecules
  • A single cell has no library to prepare: whole-genome amplification has to make one, and its evenness decides what can be called
  • Read a library by its complexity, not only by its yield

Where this connects

Blocks this protocol uses

  • Fragmentation — shearing, enzymatic cutting or tagmentation, and the size window
  • End chemistry — why end repair and dA-tailing stand between the shear and the ligation
  • Handles — adapter ligation and transposition: two of the five ways to attach a known sequence
  • Amplification — the cycle budget, GC bias, duplicates; and MDA, MALBAC, PTA for one cell
  • Barcodes and UMIs — the index that lets libraries be pooled, the UMI that separates a duplicate from a coincidence

Before and after

  • Anatomy of a library molecule — full against truncated adapters, which is what decides whether PCR is optional
  • Cleanup and QC — qPCR counts only molecules with both adapters, and the trace shows dimers and over-amplification
  • Trick catalogue — single-strand ligation for degraded input, and what to do about adapter dimers