Every library preparation exists to produce a molecule that one machine can read. A MiSeq on a bench. Photo: Konrad Förstner, CC0
So before any chemistry: what does the machine actually require, and what does it do with the molecule once it has it?
This deck is about Illumina, because it is still most of the world’s sequencing and because its constraints shaped most of the protocols you will meet. The next deck covers the others.
Sanger: the baseline
One template, one primer, one reaction — chains terminated at random by ddNTPs
Still the right tool for checking a clone, a guide site or a PCR product
Sanger et al. 1977, doi:10.1073/pnas.74.12.5463
What makes massively parallel sequencing possible
Give every fragment the same ends, and one primer can read all of them at once
The inserts stay unknown; only the ends are designed
This is why “library preparation” is a thing at all
The flow cell is a lawn of two oligos
Two sequences, P5 and P7, grafted to the surface by their 5’ ends
The object itself is the glass channel above right: a MiSeq flow cell, one lane Photo: Cirosantilli2, CC BY-SA 4.0
Your molecule must carry the complement of one at each end
Same end at both ends → no bridge → no cluster → no read
Clustering: amplify in place
The free end bends over, finds a neighbouring lawn oligo, and is copied
Repeated, this makes ~1000 identical copies in a spot a micron across
Why a cluster must be clonal
Every copy in a cluster is read at the same time, and the camera sees the sum of their signals. That has consequences you will meet constantly:
One cluster, one read. Two molecules seeding the same spot give a mixed signal and the read is filtered out.
Loading concentration matters. Too much library and clusters merge; too little and you waste the run. Getting from ng/µL to a loading molarity is QC.
Patterned flow cells (NovaSeq, NextSeq 2000) have fixed wells and use a different amplification chemistry (ExAmp: the first molecule to reach a well amplifies at constant temperature and fills it, shutting the others out), which changes how duplicates and index hopping arise.
A library is not just molecules — it is molecules at a known concentration.
One strand at a time
The cluster starts with both orientations; one kind is cut and washed away
Free 3’ ends are blocked so nothing else can prime
The sequencing primer anneals to the adapter, pointing into the insert
Sequencing by synthesis
The 3’ block is what makes it stepwise — one base per cycle, across the whole flow cell
Bentley et al. 2008, doi:10.1038/nature07517
Four colours, two, or one
How the four bases are distinguished has changed, and it affects your library:
Four-colour (MiSeq, HiSeq 2500): one dye per base
Two-colour (NextSeq, NovaSeq): red, green, both, or neither — G is “no signal”
One-colour (iSeq): distinguished across two images
The practical consequence: on two-colour instruments a long run of G is indistinguishable from a dead cluster, and a library with poly-G stretches or empty wells produces characteristic poly-G tails at the ends of reads.
Four reads, in this order
The indexes are sequence like any other, read by their own primers
Getting to the other end
Read 2 is a fresh copy of the opposite strand, not the same molecule read backwards
The i5 index is read two different ways
The commonest cause of an “empty” run that was actually fine
Dual indexing: Kircher et al. 2012, doi:10.1093/nar/gkr771
Insert size versus read length
Adapter content in a QC report is a measure of your fragment size, not your sequencing
Fix in the library: unique dual indexes (barcodes and UMIs), and clean up free adapter properly
Costello et al. 2018, doi:10.1186/s12864-018-4703-0
Base diversity
Amplicon libraries are the classic failure: every molecule starts with the same primer
Custom primers and dark cycles
Two tools worth knowing exist, because protocols in the collection rely on them:
Custom sequencing primers. If your construct does not present a standard Read 1 site where the instrument expects it, you can supply your own primer. It costs a reagent position and some goodwill from your sequencing facility, but it removes a design constraint entirely.
Dark cycles. The chemistry runs without imaging, stepping blindly past a constant stretch — a linker, a fixed barcode scaffold — so the real read starts where you want it.
Both are routine in single-cell and spatial protocols, and both are reasons a protocol’s read structure may look impossible at first glance — see the entries in the trick catalogue.
Things that degrade a run
Cluster density — too high merges clusters; too low wastes capacity
Phasing and pre-phasing — a few molecules in each cluster fall behind or run ahead, so the signal gets noisier with every cycle. This is why quality falls along a read
Free adapter — dimers cluster beautifully and index-hop
Over-amplified library — duplicates and skewed coverage, invisible until analysis (amplification)
None of these are the sequencer’s fault. All of them are set by the library.
What comes back
FASTQ per read: name, sequence, +, per-base quality
Phred score Q = −10 log₁₀(P_error): Q30 = 1 error in 1000. Modern instruments bin qualities into a few levels rather than reporting all 40
Demultiplexing splits reads by index, allowing a mismatch or two if the barcode set is far enough apart. Reads that match nothing land in Undetermined — the first place to look when a run seems empty
Read-structure notation
A run is specified as cycles per read, for example 28 / 10 / 10 / 90:
Read
Cycles
What it holds
Read 1
28
16 nt cell barcode + 12 nt UMI
Index 1 (i7)
10
sample index
Index 2 (i5)
10
sample index
Read 2
90
the cDNA insert
You design the construct; the read structure follows from it — a barcode-plus-UMI Read 1 like this one is the signature of a droplet protocol (single cells). Getting the two out of step is among the most expensive mistakes in this field, because you only find out after the run.
What this imposes on every library you build
Both ends must carry the right flow-cell sequence, and they must differ
The insert must be short enough to cluster and long enough not to be read through
Anything you want to read needs a primer site next to it, in the right orientation
Indexes must be far enough apart to survive errors, and unique in pairs if hopping matters
The first cycles must be diverse, or the run must be rescued with PhiX
The molecule must be free of adapter dimers, and quantified, before it is loaded
Where this connects
Before this
Nucleic acids — ends, hybridisation and what a ligase needs