How Illumina reads DNA

Foundations · NGS library preparation

Johan Henriksson

Design backwards from the instrument

Every library preparation exists to produce a molecule that one machine can read. A MiSeq on a bench. Photo: Konrad Förstner, CC0

So before any chemistry: what does the machine actually require, and what does it do with the molecule once it has it?

This deck is about Illumina, because it is still most of the world’s sequencing and because its constraints shaped most of the protocols you will meet. The next deck covers the others.

Sanger: the baseline

  • One template, one primer, one reaction — chains terminated at random by ddNTPs
  • Still the right tool for checking a clone, a guide site or a PCR product

Sanger et al. 1977, doi:10.1073/pnas.74.12.5463

What makes massively parallel sequencing possible

  • Give every fragment the same ends, and one primer can read all of them at once
  • The inserts stay unknown; only the ends are designed
  • This is why “library preparation” is a thing at all

The flow cell is a lawn of two oligos

  • Two sequences, P5 and P7, grafted to the surface by their 5’ ends
  • The object itself is the glass channel above right: a MiSeq flow cell, one lane Photo: Cirosantilli2, CC BY-SA 4.0
  • Your molecule must carry the complement of one at each end
  • Same end at both ends → no bridge → no cluster → no read

Clustering: amplify in place

  • The free end bends over, finds a neighbouring lawn oligo, and is copied
  • Repeated, this makes ~1000 identical copies in a spot a micron across

Why a cluster must be clonal

Every copy in a cluster is read at the same time, and the camera sees the sum of their signals. That has consequences you will meet constantly:

  • One cluster, one read. Two molecules seeding the same spot give a mixed signal and the read is filtered out.
  • Loading concentration matters. Too much library and clusters merge; too little and you waste the run. Getting from ng/µL to a loading molarity is QC.
  • Patterned flow cells (NovaSeq, NextSeq 2000) have fixed wells and use a different amplification chemistry (ExAmp: the first molecule to reach a well amplifies at constant temperature and fills it, shutting the others out), which changes how duplicates and index hopping arise.

A library is not just molecules — it is molecules at a known concentration.

One strand at a time

  • The cluster starts with both orientations; one kind is cut and washed away
  • Free 3’ ends are blocked so nothing else can prime
  • The sequencing primer anneals to the adapter, pointing into the insert

Sequencing by synthesis

  • The 3’ block is what makes it stepwise — one base per cycle, across the whole flow cell

Bentley et al. 2008, doi:10.1038/nature07517

Four colours, two, or one

How the four bases are distinguished has changed, and it affects your library:

  • Four-colour (MiSeq, HiSeq 2500): one dye per base
  • Two-colour (NextSeq, NovaSeq): red, green, both, or neither — G is “no signal”
  • One-colour (iSeq): distinguished across two images

The practical consequence: on two-colour instruments a long run of G is indistinguishable from a dead cluster, and a library with poly-G stretches or empty wells produces characteristic poly-G tails at the ends of reads.

Four reads, in this order

  • The indexes are sequence like any other, read by their own primers

Getting to the other end

  • Read 2 is a fresh copy of the opposite strand, not the same molecule read backwards

The i5 index is read two different ways

  • The commonest cause of an “empty” run that was actually fine

Dual indexing: Kircher et al. 2012, doi:10.1093/nar/gkr771

Insert size versus read length

  • Adapter content in a QC report is a measure of your fragment size, not your sequencing
  • The practical rule: size-select so the insert is comfortably longer than the read (fragmentation and size control)

Index hopping

  • Fix in the library: unique dual indexes (barcodes and UMIs), and clean up free adapter properly

Costello et al. 2018, doi:10.1186/s12864-018-4703-0

Base diversity

  • Amplicon libraries are the classic failure: every molecule starts with the same primer

Custom primers and dark cycles

Two tools worth knowing exist, because protocols in the collection rely on them:

  • Custom sequencing primers. If your construct does not present a standard Read 1 site where the instrument expects it, you can supply your own primer. It costs a reagent position and some goodwill from your sequencing facility, but it removes a design constraint entirely.
  • Dark cycles. The chemistry runs without imaging, stepping blindly past a constant stretch — a linker, a fixed barcode scaffold — so the real read starts where you want it.

Both are routine in single-cell and spatial protocols, and both are reasons a protocol’s read structure may look impossible at first glance — see the entries in the trick catalogue.

Things that degrade a run

  • Cluster density — too high merges clusters; too low wastes capacity
  • Phasing and pre-phasing — a few molecules in each cluster fall behind or run ahead, so the signal gets noisier with every cycle. This is why quality falls along a read
  • Low diversity — see above; fix with PhiX (a balanced spike-in library) or staggered primers
  • Free adapter — dimers cluster beautifully and index-hop
  • Over-amplified library — duplicates and skewed coverage, invisible until analysis (amplification)

None of these are the sequencer’s fault. All of them are set by the library.

What comes back

  • FASTQ per read: name, sequence, +, per-base quality
  • Phred score Q = −10 log₁₀(P_error): Q30 = 1 error in 1000. Modern instruments bin qualities into a few levels rather than reporting all 40
  • Demultiplexing splits reads by index, allowing a mismatch or two if the barcode set is far enough apart. Reads that match nothing land in Undetermined — the first place to look when a run seems empty

Read-structure notation

A run is specified as cycles per read, for example 28 / 10 / 10 / 90:

Read Cycles What it holds
Read 1 28 16 nt cell barcode + 12 nt UMI
Index 1 (i7) 10 sample index
Index 2 (i5) 10 sample index
Read 2 90 the cDNA insert

You design the construct; the read structure follows from it — a barcode-plus-UMI Read 1 like this one is the signature of a droplet protocol (single cells). Getting the two out of step is among the most expensive mistakes in this field, because you only find out after the run.

What this imposes on every library you build

  1. Both ends must carry the right flow-cell sequence, and they must differ
  2. The insert must be short enough to cluster and long enough not to be read through
  3. Anything you want to read needs a primer site next to it, in the right orientation
  4. Indexes must be far enough apart to survive errors, and unique in pairs if hopping matters
  5. The first cycles must be diverse, or the run must be rescued with PhiX
  6. The molecule must be free of adapter dimers, and quantified, before it is loaded

Where this connects

Before this

Read these next

Where these constraints bite

The protocols in NGS 101 are drawn with their read structure, so you can check any of them against this deck.