Quiz: what the microscope does to colocalisation

Short questions to test your understanding. Try each on paper first, then open the answer.


Q1. A constant camera offset of 100 is added to both channels. Which of Pearson’s r, the overlap coefficient and Manders’ M1/M2 (with fixed thresholds) change?

Which coefficient subtracts the means? What happens to “above threshold”?

r does not change (it subtracts the means). The overlap coefficient does (it uses raw values and moves towards 1). M1 and M2 change if the thresholds do not move with the offset: thresholds below 100 make every pixel “above”, and M1, M2 go to 1.


Q2. Uneven illumination makes both channels brighter in the centre of the field. Why does that raise r even for unrelated proteins, and how do you remove it?

Write each channel as its own signal plus a common term.

Both channels contain the same varying term, which correlates them by itself (r = k\operatorname{var}G/\sqrt{\dots}). Subtract the background before measuring: a top-hat with a disk larger than the structures, or a measured flat-field/background image, and measure in small regions where the background varies little.


Q3. Why must bleed-through be compensated before the channels are registered?

Where in channel B does the leaked light sit?

The leaked light is A’s own emission, so it sits at A’s positions, unshifted, while B’s own light carries the chromatic shift. Compensation subtracts \hat s A pixel by pixel: it removes the leaked copy only while that copy lines up with A. Registering B first moves the leaked copy away from A, and the subtraction then leaves a positive and a negative ghost of every A structure.


Q4. After compensation with \hat s = 0.6, two unrelated channels give r = -0.47. What happened?

\hat B = B + (s - \hat s)A.

The true leakage was smaller (0.3): over-compensation subtracts too much of A, leaving (s - \hat s)A = -0.3A in B, a negative copy of A. That anti-correlates the channels. Estimate \hat s from a single-stained control, and check that the compensated control channel is flat.


Q5. Two-colour beads show a shift of 0.4 pixels between the channels. Is that worth correcting for vesicles of radius 2 pixels?

Use e^{-\delta^2/4\sigma^2} with σ about 2 px.

e^{-0.16/16} = 0.99: a 1 % loss of overlap, small for r but systematic. For distance measurements between spots (unit 3) the 0.4 px (26 nm at 65 nm pixels) adds directly to every measured separation, and it is worth correcting. For a 2 px shift the overlap would drop to e^{-4/16} = 0.78.


Q6. A paper says the two proteins “interact” because their Manders coefficients are 0.9 in confocal images. What can you conclude?

How close is “colocalised” in a confocal image?

At most that the two proteins are closer than about the resolution (roughly 200 nm laterally and more axially) more often than chance, if the controls were done. “Interact” (nm distances) needs other evidence: FRET, or co-immunoprecipitation; centroid distances of well-registered single spots can get below the resolution but not to molecular contact.


Q7. (math) Show that for independent A, B of equal variance, leakage B' = B + sA gives r(A, B') = s/\sqrt{1 + s^2}. What r does 10 % leakage give?

\operatorname{cov}(A, B + sA) = s\operatorname{var}A.

\operatorname{var}(B + sA) = (1 + s^2)\operatorname{var}A, so r = s\operatorname{var}A/\sqrt{\operatorname{var}A\,(1 + s^2)\operatorname{var}A} = s/\sqrt{1 + s^2}. For s = 0.1: 0.1/\sqrt{1.01} = 0.0995, about 0.10, far outside the null spread of a block randomisation for a cell (sd ≈ 0.06), so even 10 % leakage passes as “significant”.