Quiz: correcting sensitised emission
Before this unit: FRET in pictures: a ruler for nanometres (previous unit); Day 46: Colour.
See also: Bleed-through and compensation (day 47); Day 47: Spectral unmixing; Day 16: Measurements; Day 23: FFT registration.
Short questions to test your understanding. Try each on paper first, then open the answer. I_{DD}: donor excitation, donor filter; I_{DA}: donor excitation, acceptor filter (the FRET channel); I_{AA}: acceptor excitation, acceptor filter. F_c = I_{DA} - d\,I_{DD} - a\,I_{AA}.
Q1. A cell expresses only the donor. Why is it bright in the FRET channel?
Where does the donor’s emission spectrum end?
The donor’s emission has a long red tail that reaches the acceptor filter: bleed-through, d \cdot I_{DD}.
Q2. A cell expresses only the acceptor. Why can it be (dimly) bright in the FRET channel?
Which laser is on?
The donor laser also excites the acceptor a little directly (its absorption spectrum has a blue tail): cross-excitation, a \cdot I_{AA}. The acceptor then emits in its own filter.
Q3. Which control measures d, and which a? What do you plot, and what do you fit?
In each control one of the two contaminations is absent.
Donor-only sample: I_{DA} against I_{DD}, slope through the origin = d. Acceptor-only sample: I_{DA} against I_{AA}, slope through the origin = a. Both after background subtraction, with the same settings as the experiment.
Q4. I_{DD} = 300, I_{DA} = 400, I_{AA} = 800, d = 0.5, a = 0.1, G = 2. Compute F_c and E_\text{app}.
E_\text{app} = F_c/(F_c + G\,I_{DD}).
F_c = 400 - 150 - 80 = 170. E_\text{app} = 170/(170 + 600) = 0.22.
Q5. The corrected image shows “FRET” in cells that only express free donor and free acceptor, and the brightest of those cells show the most. What is the likely cause, and how do you check it?
A residual that grows with brightness is a residual of a subtraction.
A bleed-through (or cross-excitation) coefficient that is too small: the uncorrected part is proportional to the donor (or acceptor) intensity. Check: plot F_c against I_{DD} (and against I_{AA}) in the no-FRET cells; the cloud must be flat around zero. A tilt gives the size of the error; re-measure the coefficient from the controls.
Q6. The acceptor is expressed at twice the level in a second experiment, with the same binding. Which of F_c/I_{AA}, F_c/\sqrt{I_{DD}I_{AA}} and E_\text{app} change?
F_c depends on the donors that have a partner; I_{AA} on all acceptors.
F_c/I_{AA} halves; F_c/\sqrt{I_{DD}I_{AA}} drops by 1/\sqrt2 (to 71 %); E_\text{app} stays (it only involves the donor and the sensitised emission). Only E_\text{app} can be compared between cells with different expression ratios.
Q7. Why is a pixel-wise E_\text{app} map at low FRET mostly noise, even with correct coefficients?
Variances of independent images add, also when you subtract them.
F_c is a small difference of large, noisy counts: its variance is about I_{DA} + d^2 I_{DD} + a^2 I_{AA}, set by the bright images, not by F_c. With I_{DD} = 400, I_{AA} = 1000 and F_c = 20 the noise is about 21, as large as the signal. Average over a cell or region first.
Q8. (math) Write the three images as a matrix times the three unknown signals and give the inverse matrix.
Only the middle row is mixed.
(I_{DD}, I_{DA}, I_{AA})^\top = L\,(u, F_c, v)^\top with L rows (1, 0, 0), (d, 1, a), (0, 0, 1). L^{-1} has rows (1, 0, 0), (-d, 1, -a), (0, 0, 1).
Q9. Name three practical problems that create false FRET in sensitised-emission images, besides wrong coefficients.
Offsets, geometry, concentration.
Any three of: background or offset not subtracted; the three images not aligned (edges in F_c); saturated pixels; bleaching between acquisitions; bystander FRET at high expression.