Quiz: projections and the sinogram
Before this unit: Day 1: Images, filters and frequencies.
See also: Day 4: Image formation; Day 5: Deconvolution; Day 21: Hough transform; Day 28: Ising model and Markov random fields.
Short questions to test your understanding. Try each on paper first, then open the answer. Lines are x\cos\theta + y\sin\theta = \rho, origin at the image centre, y up, \theta \in [0°, 180°).
Q1. Why does a CT scanner work with -\ln(I/I_0) rather than with the measured intensity I?
How do the attenuations of two layers combine?
Each layer removes a fraction of the photons, so the transmitted fractions multiply. The logarithm turns the product into a sum of \mu along the ray: a line integral, linear in the unknown attenuation. All reconstruction methods rely on that linearity.
Q2. 10 cm of water-like tissue (\mu = 0.2 cm⁻¹): what fraction of the photons gets through? And with 1 cm of bone (\mu = 0.5 cm⁻¹) added?
I/I_0 = e^{-\sum \mu\,\Delta s}.
e^{-2} = 0.135, about 14 %. With the bone: e^{-2.5} = 0.082; the bone alone would pass e^{-0.5} = 0.61, and the two fractions multiply: 0.135 \cdot 0.61 = 0.082.
Q3. What does the projection at \theta = 0° of an image look like, in terms of its pixels? And at \theta = 90°?
At \theta = 0° the lines are x = \rho.
At 0° the rays are vertical: the projection is the column sums, left to right with increasing \rho. At 90° the rays are horizontal: the row sums, with the top row at the largest \rho (because y points up).
Q4. A sinogram contains a sinusoid with amplitude 30 pixels whose maximum is at \theta = 60°. Where is the point?
\rho = r\cos(\theta - \varphi).
At distance r = 30 from the centre in direction \varphi = 60°: (30\cos 60°, 30\sin 60°) = (15, 26).
Q5. Why is a half turn (180°) enough for a parallel-beam scanner?
Compare the rays at \theta and at \theta + 180°.
The line at (\theta + 180°, \rho) is the line at (\theta, -\rho): the same set of points. So p_{\theta + 180°}(\rho) = p_\theta(-\rho); the second half turn repeats the first, mirrored.
Q6. The sums of the projections of a scan are 512.3 at most angles but 480.1 at a few neighbouring angles. What might have happened?
Which number should every projection share?
Every projection should sum to the total of the object. A drop at a few angles means those projections lost part of the object: for example, the object stuck out of the field of view at those angles (truncation), or detector elements failed, or the source fluctuated. The consistency check catches it.
Q7. You move an object 10 pixels up (in +y). How does its sinogram change?
Each point moves by \mathbf t\cdot(\cos\theta, \sin\theta) on the detector.
Every point’s curve moves by 10\sin\theta: not at all at 0° (vertical rays do not see a vertical shift), by +10 at 90°, and back to 0 at 180°. The whole sinogram is bent by this sinusoid; it is not simply shifted.
Q8. Row sums and column sums of a binary image are known. Why can’t you always reconstruct the image from them?
Look for a change that keeps every row and column sum.
Swapping a 2 \times 2 checkerboard pattern \begin{smallmatrix}1&0\\0&1\end{smallmatrix} \leftrightarrow \begin{smallmatrix}0&1\\1&0\end{smallmatrix} (in any two rows and two columns) leaves all row and column sums unchanged. Different images give the same two projections; more angles are needed to tell them apart.
Q9. (math) Show that every projection of f has the same integral \int p_\theta(\rho)\,d\rho.
\int p_\theta(\rho)\,d\rho = \iint f(\rho\,\mathbf n_\theta + s\,\mathbf n_\theta^\perp)\,ds\,d\rho. The map (\rho, s) \mapsto \rho\,\mathbf n_\theta + s\,\mathbf n_\theta^\perp is a rotation with Jacobian 1, so the double integral is \iint f\,dx\,dy, independent of \theta.