Quiz: resolution, sampling and measuring the PSF

Short questions to test your understanding. Try each on paper first, then open the answer. Numbers use green light, λ = 520 nm.


Q1. Two points are 0.61λ/NA apart. What does the profile through them look like? And at 0.47λ/NA?

Rayleigh and Sparrow.

At 0.61λ/NA (Rayleigh) the peak of each point lies on the first dark ring of the other: two peaks with a dip of about 26 % between them. At 0.47λ/NA (Sparrow) the dip has just vanished: the profile is flat at the midpoint, a single broad peak.


Q2. Two molecules are exactly at the Rayleigh distance, but each delivers only 20 photons. Are they resolved?

How many photons land in the brightest pixel, and how large is the Poisson noise there?

In practice no. With 65 nm pixels about 9 % of a point’s photons land in its peak pixel: about 2 photons, with noise \pm\sqrt2 \approx \pm 1.4, i.e. ±70 %. A 26 % dip is far below that noise. The Rayleigh criterion assumes a noise-free image; with few photons, noise decides what can be resolved.


Q3. A 60× / 1.4 oil objective is used with a camera of 6.5 µm pixels. Is the image Nyquist sampled at 520 nm?

Pixel in the sample plane = camera pixel ÷ magnification; Nyquist = λ/4NA.

Pixel in the sample: 6500/60 = 108 nm. Nyquist: 520/(4 \cdot 1.4) = 93 nm. Slightly undersampled (by 1.17×): the finest detail the lens passes, near the cut-off, can alias. A 1.5× extra magnifier (72 nm) would fix it.


Q4. Why not simply use pixels much smaller than Nyquist, to be safe?

What happens to the number of photons per pixel?

The optics deliver no detail beyond the cut-off, so smaller pixels add none. But the same photons are spread over more pixels: each pixel is noisier (shot noise relative to the signal grows, and every pixel adds its own read noise). Files grow and acquisitions slow down; for live samples, more light per useful pixel means more bleaching.


Q5. A widefield z-stack is taken with an NA 1.4 objective and 1 µm steps. What goes wrong?

Axial Nyquist at NA 1.4 is about 280 nm; the PSF is about 0.5 µm long (FWHM).

The z-step is about 3.6 times too large. A small object halfway between two planes (0.5 µm from each) appears in neither at full brightness: on the axis the PSF has dropped to about 4 % of its peak there. Axial detail is aliased or lost, and deconvolution cannot recover it afterwards.


Q6. What is the “missing cone” of a widefield microscope, and what does it mean for the image?

Where in 3D frequency space is the widefield OTF zero?

The 3D OTF of a widefield microscope is zero in a cone around the k_z axis: structures that vary slowly sideways but change along z (a uniform layer, a large flat sheet) are not transmitted in their z-variation. In the image, out-of-focus light from such structures is spread over neighbouring planes, not removed: widefield has no optical sectioning. A confocal pinhole fills the cone.


Q7. You measure a PSF with 200 nm beads and get an FWHM of 230 nm. Is the PSF really that wide?

\text{FWHM}_\text{meas}^2 \approx \text{FWHM}_\text{PSF}^2 + (0.53D)^2.

No: the bead itself adds about 0.53 \cdot 200 = 105 nm in quadrature. \sqrt{230^2 - 105^2} \approx 204 nm. Use smaller beads (100 nm or less) so that the correction is small.


Q8. Why average many beads instead of measuring one very carefully?

What happens to independent noise when you average N copies?

Each bead’s image is noisy; after background subtraction and sub-pixel centring, averaging N of them keeps the PSF and reduces the noise by \sqrt N (30 beads: to 18 %). A single bead can only be made less noisy with more photons, which bleaches it.


Q9. (math) Show that for a Gaussian PSF with standard deviation σ, two equal points are just at the Sparrow limit when they are 2σ apart.

The midpoint curvature of h(x - d/2) + h(x + d/2) is 2h''(d/2).

h(x) = e^{-x^2/2\sigma^2} has h''(x) = (x^2/\sigma^4 - 1/\sigma^2)\,h(x), zero at x = \sigma. The midpoint curvature 2h''(d/2) changes sign at d/2 = \sigma: below d = 2\sigma the midpoint is a maximum (one peak), above it a minimum (a dip).