Quiz: ratio imaging and the noise of a ratio
Before this unit: Acceptor photobleaching, and what an efficiency means (previous unit); Day 46: Colour.
See also: Day 47: Spectral unmixing; Day 16: Measurements; Day 23: FFT registration.
Short questions to test your understanding. Try each on paper first, then open the answer. R = I_A/I_D: acceptor over donor channel of a biosensor, both under donor excitation.
Q1. A biosensor is brighter in the nucleus than in the thin cytoplasm. Does that mean the signal (say, calcium) is higher there?
What else makes a pixel bright?
No. Brightness also depends on how much sensor is in the pixel: thickness, concentration, illumination, focus. All of these multiply both channels, so the ratio cancels them; only the ratio reports the sensor’s state.
Q2. A = 100, D = 400 after background subtraction. What ratio do you get if you forget to subtract a background of 50 in both channels?
Add 50 to each.
150/450 = 0.33 instead of 0.25: the offset pulls the ratio towards 1, more in dim pixels.
Q3. Why does a ratio image without a mask show a “carpet” of extreme colours outside the cells?
What are both channels outside the cell, after background subtraction?
Outside the cell both channels are noise around zero. The ratio of two numbers near zero is huge, negative or undefined (in the limit a Cauchy-like distribution). Mask with a threshold on the summed intensity, or use an intensity-modulated display.
Q4. A sensor has K_d = 200 nM, R_\text{min} = 0.5, R_\text{max} = 2.5. A cell shows R = 2.0. What concentration? And for R = 2.0 \pm 0.1?
c = K_d (R - R_\text{min})/(R_\text{max} - R).
c = 200\cdot1.5/0.5 = 600 nM. R = 1.9: 200\cdot1.4/0.6 = 467 nM; R = 2.1: 200\cdot1.6/0.4 = 800 nM. A 5 % error in R gives −22 %/+33 % in c, and it gets worse towards R_\text{max}.
Q5. The acceptor channel has 100 photons per pixel, the donor channel 400. What is the CV of the ratio? Which channel would you make brighter?
\mathrm{CV}^2 \approx 1/N_A + 1/N_D.
\sqrt{0.01 + 0.0025} = 0.112, 11 %. The acceptor channel: it contributes 0.01 of the 0.0125. Doubling the acceptor photons gives 8.7 %, doubling the donor photons only 10.6 %.
Q6. Over a region you can average the pixel ratios, or divide the summed channels. Which is better, and why?
Which pixels does each give the most weight?
The ratio of sums: it weights pixels by their brightness and divides two large, precise numbers. The mean of ratios gives dim, noisy pixels the same weight as bright ones and inherits the upward bias of each dim ratio (about 1/N_D).
Q7. In a time-lapse the ratio of a resting cell drifts slowly downwards. Name a cause that is not biology, and a control.
Two dyes, two bleaching rates.
The acceptor bleaches faster than the donor, so I_A/I_D falls. Control: an unstimulated cell (or the pre-stimulus baseline) imaged the same way; fit the drift and remove it.
Q8. (math) Use the delta method to show that \operatorname{E}[A/D] \approx (\mu_A/\mu_D)(1 + 1/\mu_D) for independent Poisson counts.
Expand 1/D around \mu_D to second order.
1/D = 1/(\mu_D + \delta) \approx (1/\mu_D)(1 - \delta/\mu_D + \delta^2/\mu_D^2). With A independent, \operatorname{E}[A/D] = \mu_A\operatorname{E}[1/D] \approx (\mu_A/\mu_D)(1 + \operatorname{Var}(D)/\mu_D^2) = (\mu_A/\mu_D)(1 + 1/\mu_D).
Q9. Why does a flickering lamp make each channel noisy but hardly the ratio?
What does the flicker do to the covariance of the two channels?
The flicker multiplies both channels by the same factor, which cancels in A/D. In the variance formula it appears as a positive covariance, which is subtracted.