Exercise: decode an instrument protocol

Goal

By the end of this exercise you should have:

  • decoded a complete USB recording of a (fictional) plate reader with an agent, without any hardware,
  • a protocol specification (PROTOCOL.md) where every command says how it was found,
  • a small decoder and simulated driver in Python that reproduces every recorded command byte for byte, and turns the recording back into the actions the user took,
  • checked that the checksum holds on every frame and that the decoded values match the screenshots,
  • worked through four legal scenarios (this part needs no computer and is for everyone, including leaders), and
  • an inventory of one instrument in your own lab, and a decision on what to do next.

You do not write Python. You read tables, byte listings and the agent’s claims, and you check them against what the user did. Your session will look different from anyone else’s. Compare results: the numbers and checks below.

Before you start

  • Watch or read both lectures of this track: Open control of lab instruments: why, and is it legal? and Reverse engineering instruments with agents.
  • You have done Part 2 (a project folder, git). Python 3 is installed (the agent may install it; Terminal survival kit). No extra packages are needed: Python’s built-in sqlite3 reads the recording.
  • No instrument is needed. The recording is synthetic: the LumoPlate 3000 and its protocol were invented for this course.
  • Make a folder for the track in your course folder, with a raw folder inside:
~/agentic-course/track-hardware/lumoplate/raw/

(Windows: C:\Users\<you>\agentic-course\track-hardware\lumoplate\raw\.)

The data

File What it is
lumoplate-capture.sqlite the recording: about 7 minutes of USB traffic on one bus, with the user’s clicks and screenshot notes. Save it in raw.
lumoplate-capture-events.csv the same user events as a plain table, if you want to read them in a spreadsheet.

The recording has three tables. packets: one row per USB transfer, with the time in milliseconds, the device address, the endpoint, the direction (OUT = computer to device, IN = device to computer), the transfer type and the bytes in hex. events: what the user did or saw, with the same clock. session: a few notes about the recording.

The course folder also holds make-data.py, the script that made the recording. It contains the answer. Do not copy it into your folder or show it to the agent.

What the user did (from the events table): connected, set the incubation temperature to 37.0 °C, opened and closed the door to insert a plate, shook it, read the plate at 450 nm, looked at the results, started a read at 600 nm and stopped it by mistake, set the temperature to 25.0 °C, opened and closed the door, and disconnected.

Worked example

Step 1: set up the folder and its rules

Start the agent in ~/agentic-course/track-hardware/lumoplate and type:

Make this folder a git repository with a README.md that says: this
folder decodes a synthetic USB recording of a fictional plate reader,
the LumoPlate 3000, made for a course exercise. Create the folders
scripts, notes and results. Commit.

Then the rules the agent reads every session:

Create CLAUDE.md with these rules: raw/ is read-only: never change,
move or overwrite anything in it. Use Python 3 with the standard library
only (sqlite3, struct); install nothing. Scripts go in scripts/, notes
in notes/, outputs in results/. Every claim about the protocol must be
tested against the whole recording by a script before it goes into
PROTOCOL.md, and PROTOCOL.md says for each fact how it was found and
how many frames support it.
Create AGENTS.md with these rules: raw/ is read-only: never change,
move or overwrite anything in it. Use Python 3 with the standard library
only (sqlite3, struct); install nothing. Scripts go in scripts/, notes
in notes/, outputs in results/. Every claim about the protocol must be
tested against the whole recording by a script before it goes into
PROTOCOL.md, and PROTOCOL.md says for each fact how it was found and
how many frames support it.

Check. The folder has raw/lumoplate-capture.sqlite, the three empty folders and the rules file. The .sqlite file is not changed by anything that follows: note its size now.

Step 2: have the recording described back to you

Open raw/lumoplate-capture.sqlite read-only. Describe its tables and
columns, how many packets and events there are, how long the recording
is, and which USB devices, endpoints and transfer types appear, with a
count for each. List the events table in full. Do not interpret the
bytes yet.

Check.

  • 2006 packets and 17 events; the recording runs for about 7 minutes (packet times from under 1 s to about 415 s).
  • Two device addresses: 3 and 5. Device 3 has only interrupt transfers; device 5 has control and bulk transfers.
  • The listed events match the summary under The data above. If the agent’s numbers differ, ask it to show the query it used.

Step 3: separate the instrument from the noise

Which device is most likely the plate reader, and what is the other
one? Look at transfer types, packet sizes, timing, and whether packets
appear near the user's actions. Explain the few control transfers at
the start. From now on, keep only the plate reader's bulk transfers
that carry data, and write a script scripts/extract.py that saves
them to results/frames.csv (time, direction, bytes).

Check.

  • Device 3 sends short 4-byte interrupt reports in bursts at random times: a typical USB mouse, not the instrument. Its bursts are not tied to the plate reader’s activity.
  • Device 5 is the plate reader. The six control transfers at the start are standard USB enumeration (the computer asks the device to describe itself); they happen for every USB device and are not part of the instrument’s own protocol.
  • Device 5 also has some empty bulk IN packets (length 0). The agent should drop these and say so.
  • results/frames.csv has 818 rows: 409 OUT and 409 IN. If the number is different, ask why.

Step 4: helper functions that find the patterns

Write scripts/patterns.py with small helper functions, and run them:
1. group the frames by their first 3 bytes and direction, with counts
   and lengths;
2. for each group, list the user events within 2 seconds before its
   frames;
3. find frames that repeat at a regular interval;
4. pair each OUT frame with the next IN frame and show how the first
   bytes of the two relate.
Summarise what you see in notes/patterns.md. Mark guesses as guesses.

Check. Look at the agent’s summary and ask yourself whether it fits what the user did.

  • One kind of OUT frame appears about 390 times, roughly once per second, between connecting and disconnecting, near no event in particular. That is a status poll.
  • A handful of OUT frames appear once or a few times, each just after a user action. Count them against the events: two temperature settings, four door clicks, one shake, two plate reads, one stop.
  • Every OUT frame is followed a few milliseconds later by an IN frame. The agent should notice a simple relation between the command byte of a request and of its reply. Do not accept “they seem related”; ask for the rule and how many pairs follow it.

Step 5: find the values you know

The screenshots tell you values the user chose: 37.0 and 25.0 °C, 30 s of shaking, 450 and 600 nm, and four absorbance values in the results table. Use them.

The user typed these values: temperature 37.0 and later 25.0, shaking
30 s, wavelengths 450 and 600 nm. The results table showed A1 = 2.458,
A12 = 0.048, H1 = 2.459, D6 = 0.298. Search the frames near those
events for these values in every common encoding: 8, 16 and 32-bit
integers and 32-bit floats, both byte orders. Report where each value
sits in which frame. Then decode the regular status replies in the same
way and plot or tabulate the decoded temperature over time in
results/temperature.csv.

Check.

  • Each typed value is found exactly where you would expect it, in a frame sent just after the matching click, and in one encoding that the agent names (for example “32-bit float, little-endian”). If the agent finds a value in several encodings, ask which one is consistent across all frames of that kind.
  • The four absorbance values are found in the replies the computer received just before the results screenshot (rounded to 3 decimals).
  • In results/temperature.csv the temperature starts near 22.4 °C, rises to about 37 °C within roughly a minute after the first setting, stays there, and falls to about 25 °C after the second. Open the file and look. A temperature that jumps around randomly means the wrong bytes were decoded.

Step 6: the frame layout and the checksum

Propose the general layout of a frame (start byte, command, length,
payload, and anything at the end). Then find the rule for the last
byte. Test these candidates on ALL 818 frames, with a script, and
report how many frames pass each: XOR of various byte ranges; sum of
various byte ranges modulo 256; the two's complement of that sum;
CRC-8 variants. Stop at the first rule that holds for every frame,
and say which byte range it covers.

Check.

  • The layout holds for every frame: a fixed first byte, then the command, then a length that equals the actual number of payload bytes, then the payload, then one final byte.
  • One checksum rule passes on 818 of 818 frames. A rule that passes on “most” frames is wrong. Ask the agent to show the script’s output line, not only its conclusion.
  • Ask: “Would this rule also pass if I changed one byte in a frame?” The agent should test it by flipping a byte in a copy and showing that the check fails.

Step 7: write the protocol specification

Write PROTOCOL.md: the frame layout, the checksum rule, the endpoints,
and a table of every command seen (code, name you suggest, payload
layout with types and byte order, the reply, and which user action it
belongs to). For each fact say: found how (which script), supported by
how many frames, and confidence. List what is still unknown, for
example values of the status fields that never changed in the
recording.

Check. Read it. Every command in the table should be tied to a user action you can find in the events, and every row should name its evidence. There should be an honest “unknown” section: for example, the recording shows only some values of the status byte, and only one plate type.

Step 8: a decoder and a simulated driver, tested against the log

Write scripts/lumoplate.py with:
1. decode(frame) -> a readable description, using PROTOCOL.md;
2. a LumoPlate class with methods such as set_temperature(37.0),
   door(open=True), shake(...), start_read(450), get_row(0), stop(),
   status(), each of which builds the exact bytes to send;
3. a FakeLumoPlate that answers like the recorded instrument, so the
   driver can be run without hardware.
Then write scripts/test_lumoplate.py that checks:
a) for every OUT frame in the recording, the driver method you think
   produced it builds identical bytes;
b) the checksum validates on every frame;
c) decoding the whole recording gives a timeline of actions that
   matches every event in the events table (time, action, value);
d) the decoded plate values for A1, A12, H1 and D6 match the
   screenshot.
Write the timeline to results/timeline.csv and run the tests.

Check the result.

  • The tests run and all pass, and the agent shows the output. Then ask it to break one thing on purpose (change 37.0 to 36.0 in one test) and show that the test fails. A test that cannot fail checks nothing.
  • Open results/timeline.csv next to the events table. Every event has a matching decoded action, with the right value, a few milliseconds to a few tenths of a second after the click: 37.0 °C, door open, door closed, shaking (orbital, 30 s), read at 450 nm, eight rows of results, read at 600 nm, stop, 25.0 °C, door open, door closed.
  • The decoded A1, A12, H1 and D6 match the screenshot to three decimals.
  • Two events have no command of their own: “user put a plate in the drawer” (nothing is sent) and “Disconnect” (the software simply stops polling). The agent should notice this, not invent a command.
  • raw/lumoplate-capture.sqlite has the same size as in step 1. git status shows only new files in scripts, notes, results and PROTOCOL.md.

Python or Rust? Python is fine for this exercise. For a real driver that others will rely on, the lecture recommends Rust: ask the agent to port lumoplate.py to Rust with the same tests, if you want to see what that looks like.

Step 9: a second agent checks the specification

Start the other agent (Codex if you used Claude Code, or the other way round) in the same folder, as in Parallel workers: Claude and Codex:

Read PROTOCOL.md and check every claim in it against
raw/lumoplate-capture.sqlite with your own script. Do not read
scripts/. Report claims that are wrong, unsupported, or stated more
confidently than the evidence allows.

Check. Read its report. Fix real problems with the first agent; ignore style comments. Commit.

  • Frame: A5 · command (1 byte) · length of payload (1 byte) · payload · checksum (1 byte).
  • Checksum: the two’s complement of the byte sum: (0x100 - (command + length + sum of payload)) & 0xFF. Equivalently, the bytes from the command to the checksum add up to 0 modulo 256. The start byte A5 is not included.
  • Endpoints: requests are bulk OUT on endpoint 0x02; replies bulk IN on 0x82. The reply’s command is the request’s command plus 0x80.
  • Numbers: temperatures and absorbances are 32-bit floats, little-endian; the wavelength and the shaking time are 16-bit unsigned integers, little-endian.
Request Name Payload Reply Reply payload
01 hello none 81 ASCII text LP3000 fw 2.4.1
10 status none 90 state (1 byte), temperature (float), door (1 byte: 0 closed, 1 open)
21 set temperature setpoint (float) A1 00 = OK
22 shake mode (1 byte; 2 = orbital), seconds (16-bit) A2 00
30 door 1 = open, 0 = close B0 00
40 start read wavelength in nm (16-bit), plate type (1 byte; 1 = 96-well flat) C0 00
41 get row row (0–7 = A–H) C1 row, then 12 absorbances (floats)
50 stop none D0 00
  • States seen in the status reply: 0 idle, 1 heating, 2 shaking, 3 reading, 4 data ready.
  • Noise: device 3 is a USB mouse; the six control transfers on device 5 are enumeration; empty bulk IN packets carry nothing.
  • The values: 37.0 = 00 00 14 42, 25.0 = 00 00 C8 41, 450 nm = C2 01, 600 nm = 58 02, 30 s = 1E 00. The 450 nm read is followed by eight 41 requests, one per row, starting about 24 s after Start, when the status has shown state 4.
  • No command: inserting the plate and disconnecting.

On your own project

Choose one instrument in your lab or department whose software is closed: one that annoys people, or one that holds data you worry about. You do not need to record anything. The aim is an informed decision on what to do.

Privacy first. Screenshots and logs from instrument PCs can show user names, sample or patient IDs, and network details. Do not paste them into a cloud agent unless your data rules allow it (Privacy, ethics and safety).

  1. Inventory. In your own project folder (~/agentic-course/<your-project>/), ask the agent to help you write instrument-inventory.md:

    Help me write instrument-inventory.md for the <instrument> in our
    lab. Ask me questions one at a time and record the answers: what
    the instrument is and who owns it; which computer and operating
    system controls it and whether that gets security updates; what
    connects to what (USB, serial, network) and whether it needs
    internet or a cloud account; which file formats it writes and
    whether they are documented or open; whether the vendor documents
    a protocol or offers an SDK, and on what terms; what the software
    licence and the purchase or service contract say about reverse
    engineering, data access and updates; and what we would do with
    an open driver. Do not search for or guess legal conclusions.
  2. Ask the vendor first. Ask the agent to draft a short, polite letter to the vendor asking for the protocol documentation or an SDK for interoperability, and, if the instrument is networked, how users can access the instrument’s data under the EU Data Act. Send it yourself, if you decide to.

  3. Check with your legal office before any reverse engineering: bring the inventory, the licence text and the vendor’s answer.

  4. Only then plan. If the answer is clear, ask the agent for a recording plan in the style of Reverse engineering instruments with agents (what to record, in what order, which tool, what is safe to send), as a plan only. Record nothing until the instrument’s owner has agreed.

  5. Add a line to the Log of AI use in your project README.

For leaders: the same inventory, done for the ten most important instruments in a department, is a good basis for the next procurement round.

What you should have now

  • lumoplate/ with PROTOCOL.md (with evidence for every fact), scripts/lumoplate.py and passing tests, results/timeline.csv that matches every event, and a second agent’s review, all in git.
  • Your answers to the four legal scenarios, and where your reasoning differed from the suggested answers.
  • instrument-inventory.md for one real instrument, and a decision: ask the vendor, ask the legal office, buy differently next time, or plan a recording.