Quiz: calibration and quality control

Try each question before opening its hint or answer. These practice questions check chemical reasoning and small calculations; they are not a graded question bank.

Q1. Standards set concentration and the instrument measures response. Which model direction matches this design: response ~ concentration or concentration ~ response?

Place the observed error-bearing response on the left.

Use response ~ concentration for the stated calibration model, then invert that fitted relation for an unknown. Reversing the least-squares regression is a different model, not an equivalent algebraic rearrangement.


Q2. For \(y=0.020+0.100c\) and unknown response 0.520, find c. Its preparation factor is four; what is the original concentration?

Subtract intercept, divide by slope, then undo the preparation dilution.

\(c_f=(0.520-0.020)/0.100=5.00\) mg/L final; original is 20.0 mg/L if factor four and the chemical preparation model are valid.


Q3. At a known standard, observed response is 0.310 and fitted response 0.330. What is its residual? Does that sign identify the cause?

Residual is observed minus fitted.

Residual is −0.020 response units: the observation lies below the fitted line. The sign alone does not distinguish preparation error, noise, instrument behaviour or another cause.


Q4. R² is 0.999 but residuals show a clear curved pattern. Does the high R² justify the straight line over that range?

A summary of association is not a check of every model assumption.

No. Systematic residual structure is evidence that the line misses part of the response relation. Investigate a justified range or alternative validated model rather than deleting inconvenient observations to improve appearance.


Q5. An independently sourced 5.00 mg/L check is estimated as 4.00 mg/L. The supplied acceptance window is 95–105%. What should happen?

Calculate recovery, then apply the preselected rule.

Recovery is 80%, so the affected measurement block fails this exercise rule. Withhold affected sample results, investigate, restore satisfactory calibration/checks and remeasure. Do not automatically divide all samples by 0.80 without establishing the cause.


Q6. An unknown signal is above the largest calibrated response. Can changing its recorded dilution factor without a new measurement establish its concentration?

What new evidence has been acquired?

No. A justified new dilution must actually be prepared and measured within the supported range. Editing metadata on an old reading does not supply a new response and cannot validate extrapolation.


Q7. A blank SD is 0.002 response units and sensitivity 0.250 response units per mg/L. What is the indicative 3-SD concentration increment?

Use \(3s/b\), and keep the heuristic’s scope explicit.

\(3(0.002)/0.250=0.024\) mg/L. It is an illustrative blank-relative increment, not universal detection-method validation. Procedure, matrix, decision probabilities and noise assumptions still matter; below limit is not exactly zero analyte.


Q8. Why does uncertainty in an inferred concentration include more than the SD of its repeated responses?

The inverse uses fitted slope and intercept as well as the new mean response.

Slope and intercept are estimated and correlated. Their covariance propagates through the inverse relation, along with new response variability. Preparation volumes, stock concentrations, recovery and sampling may add further contributions. A forward-response interval alone is not a concentration interval.


Return to the exercise · Lecture