Quiz: chromatographic choices and internal standards
Try the short calculations and method decisions before opening hints and explained answers. Assume all quantities refer to the stated final solution unless another basis is specified.
Q1. Which is a plausible starting method for nonvolatile heat-sensitive solutes in water: conventional GC or HPLC?
Consider entry into the mobile phase and stability under the method conditions.
HPLC is a plausible start because it transports dissolved solute without requiring ordinary GC vaporisation conditions. Solubility, stability, matrix compatibility and detector response still need evaluation; this is not a universal preference for LC.
Q2. The line is area\(=3+50c\). A resolved accepted sample has area 1003 and preparation factor 10. Calculate original concentration.
Invert the response line before applying the preparation factor.
\(c_{final}=(1003-3)/50=20\) mg/L. Original concentration is 20×10=200 mg/L, conditional on range, accepted QC, identity and preparation assumptions.
Q3. The ratio model is \(R=0.006+0.100c\). Areas are903 and 500, matched IS concentration, and preparation factor 5. What is the original caffeine estimate?
Compute903/500 first, then invert the ratio relationship.
\(R=1.806\), so \(c=(1.806-0.006)/0.100=18.00\) mg/L final. Original estimate is 90.00 mg/L. This is a point estimate under the stated calibration and evidence.
Q4. The standards all contain10 mg/L internal standard, but the sample accidentally contains 20 mg/L. Can the original ratio-versus-analyte-concentration line be applied unchanged?
What fixed design condition made that simple line appropriate?
No. The internal-standard concentration is part of the calibration relation. A different concentration changes the expected ratio. Use a justified concentration-ratio model or prepare matching samples; do not silently apply the old intercept and slope.
Q5. A standard is added after extraction. Both peaks share injection variation, but only native analyte suffered extraction loss. Does the ratio remove both effects?
Trace which steps the standard experienced.
It can compensate a sufficiently shared injection factor, but not earlier native loss that the late standard never experienced. Early addition is helpful only with appropriate equivalent behaviour, which needs evidence.
Q6. A matrix component contributes to the integrated caffeine peak but not to the internal-standard peak. Will taking their ratio restore caffeine selectivity?
The numerator contains which chemical contributions?
No. The ratio still includes the interfering contribution. Better separation or justified additional selective evidence is needed before the combined area supports caffeine quantification. A precise ratio can still be biased.
Q7. Two injections of one prepared drink agree closely. What replication claim do they support?
Distinguish collected sample, preparation and injection.
They provide evidence about repeated injection/measurement under those conditions. They are not two independent preparations or two independently obtained drinks. Shared extraction and calibration effects remain shared.