Quiz: sampling and preparation

Try each question before opening its hint or answer. These practice questions check chemical reasoning and small calculations; they are not a graded question bank.

Q1. A solid has equal-mass strata with means 5, 10 and 30 mg/kg. What is its bulk mean? What target does top-only sampling approach?

Use the material mass weights, then consider which stratum was actually collected.

The bulk mean is \((5+10+30)/3=15\) mg/kg. Top-only grabs approach the top mean of 5 mg/kg under this model. Smaller top-only scatter does not supply information about the rest of the solid.


Q2. The same strata instead have mass fractions 0.20, 0.30 and 0.50. What is the bulk mean?

Multiply each concentration by its corresponding mass fraction.

\(0.20(5)+0.30(10)+0.50(30)=19\) mg/kg. An equal-stratum average would target three equal masses and is no longer appropriate. The weights must refer to the defined material, not convenience.


Q3. A 1.000 g material gives 50.00 mL extract. A 5.00 mL aliquot made to 25.00 mL reads 2.00 mg/L. Find extracted analyte in mg/g.

Undo the aliquot dilution, find total extracted mass, then divide by original mass.

The extract is \(2.00(25/5)=10.0\) mg/L. It contains \(10.0(0.05000)=0.500\) mg; divided by 1.000 g gives 0.500 mg/g extracted. Complete native recovery still needs evidence.


Q4. With D=4, 20 mL aqueous and 5 mL extracting phase, what is the ideal recovered fraction?

Use \(DV_o/(V_a+DV_o)\).

\(4(5)/(20+4(5))=0.50\), or 50%. D is a concentration ratio; recovery also depends on phase volumes. This assumes equilibrium, immiscibility, unchanged volumes and no other loss.


Q5. Why can two fresh solvent extractions recover more than one extraction using the same total solvent?

After removing each extracting phase, the next fresh portion repartitions the analyte left behind.

Each fresh portion establishes a new equilibrium with the remaining analyte. The fractions left after each step multiply. This can improve ideal recovery without adding more total solvent, but does not prove real recovery if equilibration or separation fails.


Q6. A post-extraction spike recovers 100%. What claim about native extraction remains unsupported?

Identify the steps the added analyte skipped.

Complete recovery of native analyte before spike addition remains unsupported. The spike did not undergo that extraction, and even a pre-extraction freely dissolved spike may not represent native bound material.


Q7. Does one composite made from ten increments count as ten independent samples for estimating spatial variation?

What happened to the original increment identities and separate measurements?

No. The composite is one combined material under its specified weighting design. It can support a bulk quantity, but it does not preserve ten independent measured spatial results or their original variation.


Q8. Why might a treatment useful for retaining an element in solution be unsuitable for preserving its original oxidation states?

Different measurands require different chemical information to survive.

A preservation treatment can prevent precipitation or surface loss while changing speciation. The result must match the intended elemental amount or species quantity and the specified preparation. There is no universally suitable preservation treatment for every measurand.


Return to the exercise · Lecture