Quiz: error and uncertainty

Try each question before opening its hint or answer. These practice questions check chemical reasoning and small calculations; they are not a graded question bank.

Q1. Method A gives repeated values near 10.0 mg/L with SD 0.30; B gives values near 10.5 with SD 0.02. The reference is 10.0. Which is more precise, and which shows a positive mean difference?

Treat spread and reference difference as separate properties.

B is more precise under these repeat conditions but has an observed positive difference of about 0.5 mg/L from the reference. Good repeatability is not the same as low bias. A rigorous bias assessment would also consider reference uncertainty and design.


Q2. Nine independent stable readings have sample SD 0.30 mg/L. What is the reading contribution to uncertainty in their mean?

Use \(s/\sqrt n\) for this specific contribution.

\(0.30/\sqrt9=0.10\) mg/L. Individual readings still have spread about 0.30 mg/L. The smaller mean uncertainty does not include any shared stock, preparation or sampling component.


Q3. A mean has an independent reading contribution 0.03 mg/L and a shared component 0.04 mg/L. Find combined standard uncertainty.

Add squared independent contributions, then take the square root.

\(u=\sqrt{0.03^2+0.04^2}=0.05\) mg/L. This is a standard uncertainty, not automatically a 95% interval or a complete method budget.


Q4. What happens to the previous combined uncertainty if indefinitely many independent readings drive the reading component towards zero?

Which component is shared by the readings?

It approaches 0.04 mg/L, not zero. More readings of the same solution do not independently re-establish the shared component.


Q5. Every reading has the same known +0.20 mg/L bias. Will increasing n remove it?

A fixed common displacement is not independent zero-mean noise.

No. The mean retains the fixed bias. Investigate or use an appropriately justified correction with its uncertainty. An arbitrary larger or smaller error bar does not itself correct the result.


Q6. Two volume uncertainties arise partly from the same temperature error. Can they always be combined as independent squared terms?

Shared causes can induce covariance.

No. Covariance terms can matter. The independent-input square-sum formula is conditional on independence; the signs of model sensitivities and the correlations determine how shared effects propagate.


Q7. A preparation has repeated identical displayed readings, so its observed SD is zero. Is the method uncertainty zero?

Which sources can move every reading together without changing their scatter?

No. Common calibration, preparation, recovery, display resolution and sampling effects can remain. Repeating one prepared solution does not reveal every relevant source, and zero displayed scatter is not a validation of exactness.


Q8. A result is reported with standard uncertainty 0.05 mg/L. Can the reader interpret ±0.05 as a 95% interval?

A coverage interpretation needs a construction and assumptions.

Not from that statement alone. Standard uncertainty is on an SD-like scale. Expanded uncertainty needs a stated coverage factor and rationale; a confidence interval needs a sampling/model construction and level. State which quantity and components are included.


Return to the exercise · Lecture