Quiz: analytical evidence and reporting decisions

Each answer should connect a number to its unit, sample basis and supporting evidence. A calculable candidate is not automatically a reportable chemical result.

Q1. A drink assay reports a peak area. State a suitable intended measurand for caffeine quantification.

Include analyte, material and quantity/unit rather than merely naming an instrument.

For example: caffeine mass concentration in the supplied drink, expressed in mg caffeine/L, under the defined preparation and sampling scope. Peak area is the response used to infer that quantity, not the measurand itself.


Q2. The calibration is area\(=4+40c\), valid over 0–30 mg/L final. An old sample reading gives 40 mg/L final. May you validate it just by editing the dilution factor?

Would that create a new response from a new prepared solution?

No. The measured solution was outside range. Prepare and measure an appropriate new dilution and record its actual aliquot and final volume. An edited factor does not add new measurement evidence.


Q3. A new accepted4.00 mL aliquot made to40.00 mL gives area 804. Calculate original concentration.

Invert the response model, then use40/4.

Final concentration is(804−4)/40=20 mg/L. The preparation factor is 10, giving200 mg/L original, conditional on the accepted identity, selectivity, QC and preparation evidence.


Q4. An unknown is within calibration range, but an independent check for its block recovers85% under a95–105% exercise rule. What should happen to the result?

Range and independent QC ask different questions.

Withhold the affected quantitative result and investigate/re-establish satisfactory calibration and checks before remeasurement. Being in range does not override failed QC. The failure alone does not prove a particular cause or justify dividing every value by 0.85.


Q5. The final estimate has standard uncertainty 0.20 mg/L and dilution factor 25. What is the dilution-scaled standard uncertainty on the original basis? Is it a 95% interval?

Propagate the supplied final estimate through the fixed scale factor; identify omitted inputs.

It is 0.20×25=5.0 mg/L under a fixed-factor model. It is a standard uncertainty, not automatically a 95% half-width. Actual dilution-volume, sampling, recovery and other omitted uncertainties would need evaluation for a full budget.


Q6. An ideal rounded calibration fits exactly. Does that establish zero uncertainty in real sample concentrations?

Which sources are absent from those ideal teaching points?

No. Real reading, standard preparation, volume, recovery and matrix effects can remain. Zero residual in deliberately idealised points is not a full uncertainty study; the final case supplies separate uncertainty inputs explicitly.


Q7. Two methods correlate extremely well, but one has increasing positive differences from a suitable reference. Are they interchangeable?

Association and agreement are different questions.

Not on correlation alone. Examine differences with reference uncertainty, sampling scope and the required analytical performance. A biased method can preserve rank and correlation while giving unsuitable amounts. The pattern motivates investigation rather than proving a chemical mechanism.


Q8. An ion near m/z195 is observed. Does that alone uniquely establish caffeine identity or its concentration?

Consider charge, protonation, fragments and calibration separately.

No. Singly protonated caffeine can appear near 195 even though its neutral nominal mass is 194, but other ions can share nominal mass. Suitable retention, standards, ion/fragment information and blanks can strengthen identity. Quantification separately requires an appropriate calibrated response and sample-preparation model.

Lecture · Exercise and embedded exploration