Quiz: matrix effects and standard addition
Try each question before opening its hint. Then explain the answer in chemical and measurement terms. Lecture · Exercise.
Q1. Why do the standard-addition flasks use equal sample aliquots and equal final volumes?
Track native concentration and matrix fraction.
Both remain constant across flasks, while added-analyte concentration varies. Changing final volume without correction would also change native concentration and matrix fraction, invalidating the simple fixed-volume intercept formula.
Q2. Matched unspiked/spiked final solutions give external estimates 2.00 and 2.80 mg/L after adding 1.00 mg/L. Calculate incremental recovery.
Use the change in the estimate, not the full spiked value.
Recovery is \(100(2.80-2.00)/1.00=\mathbf{80\%}\). It tests incremental response from the point of addition under the matched design, not automatically total native recovery.
Q3. Equal 4.00 mL sample aliquots made to 20.00 mL give \(y=0.360+0.120c_{added,final}\). What original concentration is inferred?
First find the negative x-intercept; then undo sample dilution.
\(x_0=-0.360/0.120=-3.00\) mg/L final. Its magnitude multiplied by \(20/4=5\) gives 15.0 mg/L original under the stated additive model. Negative concentration was not physically prepared.
Q4. An unrelated constant signal remains in every flask. Can an exact straight standard-addition line prove the native analyte estimate is correct?
What else contributes to the intercept?
No. A constant additive interference increases the intercept and is misread as native analyte. Shared multiplicative sensitivity cancels in intercept/slope, but an uncorrected additive term does not. Independent background/selectivity evidence is needed.
Q5. A spike added after extraction recovers 100%. Can native analyte still have been lost during extraction?
Ask which steps the added material experienced.
Yes. The spike bypassed the earlier loss and tests only later behaviour. Even a pre-extraction freely dissolved spike may not mimic native bound analyte; the chemical equivalence assumption needs evidence.
Q6. Does an ordinary response confidence interval for the fitted intercept automatically give a confidence interval for original standard-addition concentration?
The target quantity is a ratio of coefficients times a volume ratio.
No. The target depends on intercept, slope, their covariance and preparation volumes. A forward response interval has another target; a suitable inverse uncertainty calculation and its assumptions are needed.